XD 10, XD
⁺
10, XD 14, XD
⁺
14, XD 17 and XD
⁺
17 Exposure
Sequence
Topics:
•
•
•
Modality Settings
Apply these modality settings:
Table 3: Modality settings
kV
75 kV
mAs (indica-
tive value for
first exposure
level)
25 mAs (XD 10, XD
⁺
10)
32 mAs (XD 14, XD
⁺
14, XD 17, XD
⁺
17)
mAs (indica-
tive value for
second expo-
sure level)
12.5 mAs (XD 10, XD
⁺
10)
4 mAs (XD 14, XD
⁺
14, XD 17, XD
⁺
17)
Focus
large focus
SID
Use the (average) SID for normal operation.
Position
If the DR Detector is used in a fixed position, use the same
position for calibration.
If the DR Detector is used in different positions, perform
the calibration with the short side of the DR Detector paral-
lel to the axis of the X-ray tube.
DR Detector Calibration | Calibrating the DR Detector |
17
0134I EN 20210924 1204