
DX-D 40C and DX-D 40G Exposure Sequence
Topics:
•
•
Modality Settings
Apply these modality settings:
Table 10: Modality settings
kV
75 kV
mAs (indicative value for
first exposure level)
25 mAs
mAs (indicative value for
second exposure level)
12.5 mAs
Focus
large focus
SID
Use the (average) SID for normal operation.
Position
If the DR Detector is used in a fixed position, use
the same position for calibration.
If the DR Detector is used in different positions,
perform the calibration with the short side of the
DR Detector parallel to the axis of the X-ray tube.
Grid
Remove the grid.
Filtration
Mount the Agfa calibration filter (Cu 1.5 mm).
Check that there is no additional filtration.
DR Detector Calibration | Calibrating the DR Detector |
31
0134I EN 20210924 1204