 Getting started with DR 100e - Performing an exposure 

0366A EN 20170421 

Selecting the exposure 

In NX, select the thumbnail for the exposure in the Image Overview 

pane of the Examination window.  

The DR Detector is activated. The DR Detector Switch shows which 
DR Detector is active and shows its status.  

The default X-ray exposure parameters for the selected exposure are 
sent to the modality and displayed on the X-ray generator control 

panel. The control panel can be used to set different exposure 
parameters. 

    

    

 

 

 

Preparing the exposure 

 

Check the position of the X-ray unit and the patient. 

 

WARNING: 

Liquids ingressing the DR Detector may cause 

malfunction and contamination. 

 

If there is a chance that the detector comes in contact 

with liquids (bodily fluids, disinfectants,...), the DR 
Detector must be wrapped in a protective plastic bag 
while performing the examination. 

 

Switch on the light localizer on the collimator. 

 

If required, adapt collimation. 

 

Checking the exposure settings 

On the NX application: 

 

Check if the DR Detector Switch displays the name of the DR 

Detector that is being used.  

 

If a wrong DR Detector is displayed, select the right DR Detector 

by clicking the drop down arrow on the DR Detector Switch. 

On a DR Detector that has a status indicator: 

 

Check if the status of the DR Detector is ready for exposure. If 

the status is not ready for exposure, the DR Detector cannot be used 

for making an exposure. 

On the X-ray generator control panel: 

 

Check if the exposure settings displayed on the console are 

suitable for the exposure. If other exposure values are required than 
those defined in the NX exam, use the control panel to overwrite 

the default defined exposure settings. 

On the control panel: 

 

Check if the led beacon light lights up green. This indicates that 

the wireless DR Detector and the X-ray generator are ready for 
making an exposure. 

 

Executing the exposure 

 

Activate the preferred exposure control mode. 

 

Switch between remote control exposure button or wired 

exposure button. The indicator lights up when the remote 

control is enabled. 

 

WARNING: 

Excessive user or operator radiation exposure. Always 
keep a distance of at least 2 meters from the focal 
spot and X-ray beam, protect body and do not expose 

hands, arms or other parts of the body to the primary 
beam. 

 

Execute the exposure. 

 

WARNING: 

 

The radiation indicator on the control console 

lights up during exposure release. The led beacon 
light lights up yellow. 

 

CAUTION: 

Do not select another thumbnail until the preview 
image is visible in the active thumbnail. The acquired 

image may be linked to the wrong exposure. 

 

In NX: 

 

The image is acquired from the DR Detector and displayed in the 

thumbnail. 

 

If collimation is applied, the image is automatically cropped at 

the collimation borders. 

 

The actual X-ray exposure parameters are shown in the Image 

Detail pane 

10 

Perform quality control 

 

Note: If you use a wired network, NX can only send 

images if the network cable is plugged in. 

In NX: 

 

Select the image on which quality control is to be performed. 

 

Prepare the image for diagnosis by using e.g. L/R markers or 

annotations. 

 

If the image is OK, send the image to a hardcopy printer and/or 

PACS (Picture Archiving and Communication System). 

11 

Stopping the device 

 

Bring the device to transport position. 

 

Press the power off button. 

 

 

Unplug the power supply cable. 

 

Wind the mains cable on the cable reel. 

 

Move the device to a suitable place for parking. 

 

Switch off the DR Detector and charge the battery. 

In NX: 

 

Go to the Windows Start menu and select Shut down. 

 