6.
Clean the image plate as follows:
• Take an appropriate wipe or moisten the cloth with cleaner and wipe
the
image plate surface
softly and evenly.
Warning
: If CR Phosphor Plate Cleaner is used, do
not pour it directly onto the image plate.
• Leave the
image plate surface
to dry for about 10 minutes to allow the
solvents to evaporate.
• Once the image plate surface is dry (after about 10 minutes), check
once again for particles of material and other impurities before placing
the image plate into the cassette.
7.
Put the image plate back into the cassette.
Verify that the white phosphor side is oriented to the black tube side of the
cassette and that the shutter does not scratch the image plate.
4
| Cleaning the image plate of CR HD5.0, CR MD4.0R, CR HM5.0 and CR MM3.0R | Cleaning the
Image Plate
0042A EN 20170718 1541