47
4454C EN 2013-05-14
Reading an image plate
CR 35-X D
IGITIZER
The CR 35-X converts the information of the latent image to digital data.
4
Enter the identification data via the ID Station.
For detailed information, refer to the User manual of the ID Software.
As soon as you have entered the identification data, the CR 35-X displays:
As soon as the CR 35-X has digitized the entire image plate and you have entered the
identification data:
• the CR 35-X erases the image plate and re-inserts it into the cassette;
• the CR 35-X gives the cassette ID data the status ‘erased’.
• the cassette is returned to the cassette slot;
• the digital image data is sent to the image processing station (‘destination’).
When the CR 35-X has treated the cassette, it displays the operator main screen.
5
Remove the cassette from the cassette slot.
Note: If the CR 35-X displays an error message, refer to
operation (‘Key-operator mode’)’
WARNING:
When the CR 35-X returns the cassette, it is ready to be re-
used immediately. However, if you leave it for more than 3 days before
re-using it, you must re-erase it first. Refer to
on page
.
BUSY
DIRECT ID
Identify at ID station
or press to use emergency data
BUSY
Miller Chest
AP
Summary of Contents for CR 35-X
Page 1: ...CR 35 X User Manual 4454C EN 2013 05 14 ...
Page 4: ...4 4454C EN 2013 05 14 CR 35 X DIGITIZER ...
Page 40: ...40 4454C EN 2013 05 14 Switching off the CR 35 X CR 35 X DIGITIZER ...
Page 70: ...70 4454C EN 2013 05 14 Troubleshooting checklist CR 35 X DIGITIZER ...
Page 71: ...Equipment information sheet Appendix A ...
Page 77: ...CR cassettes Appendix B ...
Page 88: ...88 4454C EN 2013 05 14 Technical specifications of the image plates CR 35 X DIGITIZER ...
Page 89: ...Remarks for HF emission and immunity Appendix C ...
Page 95: ...95 4454C EN 2013 05 14 Remarks for HF emission and immunity CR 35 X DIGITIZER ...