• The red dot must be situated under the anode side. When the
anode is on the left side, the red dot must be at the lower left hand
side. When the anode is on the right side, the red dot must be at
the upper right hand side. In most situations the anode will be at
the left hand side, the cathode right.
• Make sure the light lines match the according lines and arrows on
the positioning template.
Cassette / Phanthom
PO
SIT
ION
ING
TEM
PLA
TE
AG
FA
Figure 8: Position the positioning template.
c) As the positioning lines are asymmetrical, align your light field so that
on the cathode side the light touches the end of the arrow and the
complete positioning template is covered. On the anode side the light
field will exceed the positioning template.
d) Make sure that the Heel effect is perpendicular with the longest side of
the positioning template.
2.
Position the cassette on the positioning template.
a) Put the cassette on the template as shown below:
1.
The scribor or the label of the detector / cassette must be at the
bottom left.
2.
For other types of cassettes (MD10, MD 4.0,...), the label of the
cassette must be at the bottom left.
3.
If you use small cassettes: the cassette must be pushed against the
metal pins in the positioning template.
Auto QC² | Auto QC² Hardware Components |
29
2370C EN 20180330 1524