Preventive maintenance schedule
The printer is designed for trouble-free operation. Maintenance and cleaning
involve only some minor user tasks. Refer to the following pages for the
appropriate cleaning procedure.
Interval
What to do?
Ad hoc
‘Cleaning and disinfection’
Ad hoc
‘Print head cleaning’
Ad hoc
‘Cleaning the dust rollers’
Each 6 months
(or more if required).
‘Cleaning the cooling air flow holes’
When image quality tends to degrade.
‘Print head profile calibration’
Safety guidelines
WARNING:
To prevent damage to the printer while performing
maintenance, observe the following safety precautions:
• Do not lubricate the printer.
• Do not attempt to disassemble the printer.
• Do not touch the resistor line of the print head.
• Always switch off the printer and disconnect the power cord
from the outlet before carrying out any maintenance work.
Note
:
Film jam removal or cleaning the printer head can be done
without switching the power off.
Recurrent safety tests
The printer shall be tested according to IEC 62353* in a time interval of at
least 36 months or less if local regulations are different.
*Medical electrical equipment – Recurrent test and test after repair of medical
electrical equipment.
DRY 3.4 | Preventive maintenance and replacements |
111
2602A EN 20191018 0826
Summary of Contents for 5364/500
Page 1: ...DRY 3 4 5364 500 User Manual 2602A EN 20191018 0826 ...
Page 12: ...12 DRY 3 4 DRY 3 4 2602A EN 20191018 0826 ...
Page 58: ... Audio signals The keypad The display 58 DRY 3 4 Introduction 2602A EN 20191018 0826 ...
Page 87: ...The menu structure DRY 3 4 Advanced Operation key operator mode 87 2602A EN 20191018 0826 ...