Chapter 4 Maintenance Procedures
95
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
To drain and fill the ultrasonic wash supply bottle:
WARNING
!
WARNING!
You will be handling ethanol in this procedure. For safe handling procedures, refer
to the MSDS provided by your ethanol supplier. At a minimum: 1) wear protective clothing
(laboratory gloves, eye protection, and laboratory coat), 2) have adequate laboratory ventilation,
3) keep away from ignition sources, 4) and have proper fire extinguishing devices/solutions readily
accessible.
1.
Prepare at least 100 mL of a 50:50 solution by volume of Type 1, > 18.2
M
Ω
water and
100% ethanol.
2.
Go to the Maintenance screen (tap the Back button
on any screen and tap the
MAINTENANCE
button
on the Main Menu).
3.
Tap
SONICATOR SOLUTION Drain
on the Maintenance screen
This drains the ultrasonic wash. An on-screen message displays the time remaining
for the operation. Wait for draining to complete.
4.
Tap
SONICATOR SOLUTION Fill
.
This moves the pin array head so that you can access the ultrasonic wash supply
bottle.
Figure 4.8 Draining the Ultrasonic Wash
Figure 4.9 Maintenance Screen