Chapter 2 Dispensing to SpectroCHIP Arrays
38
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
2.
Under
calibrant
, tap the
dispense speed
value. Enter a dispense speed using the
numeric keypad that appears, and tap
enter
.
• For iPLEX and UltraSEEK, enter 100.
• For MassCLEAVE, enter 140.
To select analyte and/or calibrant dispensing:
Analyte is dispensed to the main pads on a SpectroCHIP Array. Calibrant is dispensed to
the calibrant pads on the SpectroCHIP Array.
Select an option under
operation
Figure 2.24 Set the Calibrant Dispense Speed
Figure 2.25 384 SpectroCHIP Array
Figure 2.26 Dispensing Options
Tap and enter a
dispense speed value.
$*(1$
10 calibrant pads on a 384 SpectroCHIP Array
(5 calibrant pads on a 96 SpectroCHIP Array)
Main pads
analyte only
- Dispenses only analyte from
microtiter plates to SpectroCHIP Arrays.
calibrant only
- Dispenses only calibrant from
the calibrant reservoir to the calibrant pads
on the SpectroCHIP Arrays.
analyte and calibrant
- Dispenses both
analyte and calibrant to the SpectroCHIP
Arrays.