Appendix C External Tanks
149
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
6.
Close the front access panel and the main door.
7.
Tap
OK
in the on-screen instructions
.
The water from the internal supply tank is pumped through the wash system and
empties into the internal waste tank which, in turn, empties into the external waste
tank.
Fill the internal supply tank half-full:
1.
Manually fill the internal supply tank to half capacity (see
for instructions).
2.
Manually prime the water supply system (see
for instructions).
3.
Reattach the tank access panel.
Return the supply valves to the external supply setting:
1.
Open the main door and lift the front access panel to access the tank control valves.
2.
Set the supply valves to position "A1" (see
).
3.
Close the front access panel and the main door.
4.
Tap the
HOME
button
in the “safety interlock is disengaged” on-screen message.
5.
Fill the external supply tank with Type 1, > 18.2
M
Ω
water.
6.
Reinsert the supply tubing into the external supply tank.
Make sure that the end of the tubing with the metal fitting is at the bottom of the tank
so that air will not be drawn into the tubing. The metal fitting helps compensate for
the tendency of the tubing to float when it is empty.
Figure C.16 Draining the Supply Tank
This message disappears when draining is done.