Appendix C External Tanks
143
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
C.6 Cleaning and flushing the wash system (for external tanks only)
NOTE
These instructions are for a setup using an external supply tank and an external waste tank. If you
are using a combination of internal and external tanks, for example, an internal supply tank with
an external waste tank (or vice versa), please contact Agena Bioscience Customer Support for
assistance.
This section explains how to clean the supply and waste tanks with ethanol and then
flush the wash system with a 65% ethanol solution.
CAUTION
!
It is critical that the cleaning and flushing procedures be performed monthly to prevent microbial
growth in the wash system.
WARNING
!
WARNING!
You will be handling ethanol in this procedure. For safe handling procedures, refer
to the ethanol MSDS provided by your supplier. At a minimum: 1) wear protective clothing
(laboratory gloves, eye protection, and laboratory coat), 2) have adequate laboratory ventilation,
3) keep away from ignition sources, 4) and have proper fire extinguishing devices/solutions readily
accessible.
Drain the external
waste tank
1.
Empty the external waste tank by following the instructions on
.
Drain the internal
supply tank
1.
Go to the Maintenance screen (tap the
Back
button
on any screen and tap the
MAINTENANCE
button
on the Main Menu).
2.
Tap
SUPPLY TANK Drain
on the Maintenance screen.
Instructions on valve operation appear
3.
Open the main door and lift the front access panel to access the tank control valves.
(See
for instructions on opening the front access panel.)
Figure C.9 Draining the Supply Tank