Appendix A Changing the Pin Format
119
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
c. Leave the main door open and tap
OK
.
.
NOTE
Your pin protection block may look different that the one in the photos below.
4.
Place the pin protection block onto the dry station
.
Make sure that the pin protection block alignment pins fit into the correct dry station
well
Figure A.4 Remove the Vacuum Cover and Calibrant Reservoir
This image shows the 6-pin vacuum cover. If your instrument is currently adapted to the 1-pin format,
the 1-pin cover will be installed. It looks just like the 6-pin cover except it has only one hole in it.
Figure A.5 Instructions for Next Steps
Calibrant
reservoir
Vacuum cover
If you are currently using the
standard 24-pin format, no
vacuum cover will be present
on the dry station. In this
case, you need only remove
the calibrant reservoir.