Chapter 1 Introduction
9
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
Main door and front
access panel
CAUTION
!
Opening the main door disengages the safety interlock and aborts the dispensing run in progress.
for more details on the enclosure safety interlocks.
Opening or closing the main door
To open the main door, lift the door fully open using the handle
. Once the
door is approximately half-way open, the gas springs will assist lifting the door. The door
remains in the open position until you close it. The safety interlock message is displayed
while the door is open
To close the main door, grasp and hold the handle as you close the door. Tap the
HOME
button in the Safety Interlock message after the door is closed.
WARNING
!
WARNING!
Do not place hands in the main door pathway when closing it. When lowering the
main door, do not let go of the handle until the door is completely closed. The gas springs support
the door as it is lowered, but stop providing support when the door is almost closed. If you do not
hold onto the door handle, the door may slam closed.
Figure 1.8 Nanodispenser RS1000 front side
Handle