Chapter 4 Maintenance Procedures
104
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
3.
After tank is drained and the flow of waste fluid from the tank stops, disconnect the
tubing from the Waste port. To disconnect the tubing, press down on the release tab
and pull the quick-disconnect fitting out
.
NOTE
A small amount of residual water may be in the tubing. When the tubing is pulled out of the waste
port, hold the end of the tubing up so that residual water drains out the other end back into the
waste reservoir or sink.
4.5 Monthly maintenance
This section explains how to clean the supply and waste tanks with ethanol and then
flush the wash system with a 65% ethanol solution.
Figure 4.22 Connecting Waste Tubing to the Waste Port
The left side of the instrument is shown.
Figure 4.23 Disconnecting the Waste Tubing From the Waste Port
Waste port
Press