AGA Medical AMPLATZER TorqVue Instructions For Use Download Page 3

3

Directions for Use

Materials recommended for use with the delivery system

Exchange-length 0.035-inch guidewire

Procedure

CAUTION: Refer to the instructions for use provided with the device when placing an AMPLATZER device using an 
AMPLATZER TorqVue Delivery System.

General instructions for the AMPLATZER TorqVue Delivery Systems are provided below.

 1.  Select the appropriate delivery system for use with the device being delivered. Refer to Table 1 and Table 2 to select a 

delivery system by AMPLATZER device size, or refer to Table 3 to select the delivery system with a sheath that has the 
correct inner diameter as indicated in the device’s instructions for use.

2. Prepare the delivery system for use:

- Inspect the sterile pouch and verify that it is unopened and undamaged. Do not use the components if the sterile 

barrier has been compromised.

- Gently open the sterile pouch and inspect the components for damage. Do not use damaged or kinked components.
- Flush all components with sterile saline.
- Wipe the dilator and sheath with sterile gauze moistened with sterile saline to remove any foreign material.

3. Access the desired vessel.
4. Place the guidewire according to the device’s instructions for use. 
5. Insert the dilator into the sheath. You may encounter resistance as the dilator reaches the distal end of the sheath 

because the end of the sheath is tapered. 

6. Turn the rotating luer on the dilator clockwise to lock the components together.
7. Advance the dilator and sheath over the guidewire until the sheath is positioned according to the device’s instructions for 

use.

8. Turn the rotating luer on the dilator counterclockwise to unlock the dilator. Slowly remove the dilator from the sheath. 

WARNING: Remove the dilator slowly to prevent an ingress of air.

9. Allow blood backflow to purge any air from the sheath.

10. Attach the hemostasis valve to the proximal end of the loader and flush with sterile saline.

CAUTION: Use the hemostasis valve to impede blood backflow during the implant procedure.

11. Capture the device in the loader according to the device's instructions for use.
12. Attach the distal end of the loader to the proximal end of the sheath. Turn the rotating luer on the loader clockwise to lock 

the components together. Remember to close the stopcock on the hemostasis valve.

AMPLATZER Muscular VSD Occluder

4 mm

6–10 mm

12 mm

14 mm

16 mm

18 mm

Table 3. Delivery System Dimensions

Delivery system 

(sheath) size

Inner diameter of 

sheath

Outer diameter of 

sheath

5 Fr

1.83 mm (0.07 in)

2.51 mm (0.10 in)

6 Fr

2.11 mm (0.08 in)

2.79 mm (0.11 in)

7 Fr

2.44 mm (0.10 in)

3.18 mm (0.13 in)

8 Fr

2.69 mm (0.11 in)

3.45 mm (0.14 in)

9 Fr

3.00 mm (0.12 in)

3.81 mm (0.15 in)

10 Fr

3.30 mm (0.13 in)

4.14 mm (0.16 in)

12 Fr

3.99 mm (0.16 in)

4.80 mm (0.19 in)

Table 2. Compatibility Chart for TorqVue 180° Delivery System and AMPLATZER Devices

180° Delivery System Sizes

5 Fr

6 Fr

7 Fr

8 Fr

9 Fr

Summary of Contents for AMPLATZER TorqVue

Page 1: ...le from a device Indications and Usage The AMPLATZER TorqVue delivery system is intended to provide a pathway through which devices are introduced within the chambers and coronary vasculature of the heart or in the peripheral vasculature Contraindications None known Warnings Use on or before the last day of the expiration month noted on the product packaging The delivery system is sterilized using...

Page 2: ...s Potential Adverse Events Potential adverse events that may occur during or after a procedure using this delivery system may include but are not limited to Device Compatibility The AMPLATZER devices compatible with the TorqVue delivery system are identified in Table 1 and Table 2 CAUTION No devices other than those listed in these tables have been tested for use with the delivery system Using unt...

Page 3: ...nd of the sheath is tapered 6 Turn the rotating luer on the dilator clockwise to lock the components together 7 Advance the dilator and sheath over the guidewire until the sheath is positioned according to the device s instructions for use 8 Turn the rotating luer on the dilator counterclockwise to unlock the dilator Slowly remove the dilator from the sheath WARNING Remove the dilator slowly to pr...

Page 4: ...ligation under this warranty is limited to replacing or repairing at its option at its factory this product if returned within the warranty period to AGA Medical Corporation and after confirmed to be defective by the manufacturer EXCEPT AS EXPRESSLY PROVIDED IN THIS WARRANTY AGA MEDICAL CORPORATION DISCLAIMS ANY REPRESENTATION OR WARRANTY OF ANY KIND EXPRESS OR IMPLIED INCLUDING ANY WARRANTY AS TO...

Page 5: ...ecommended delivery sheath catheter dimensions Federal law USA restricts this device to sale by or on the order of a physician or properly licensed practitioner Indication of conformity with the essential health and safety requirements set out in European Directives ...

Reviews: