APPENDIX C. Parameters available for editing in Monitor
BUF (310) application
Table C.1. System tab (Fig.9)
Parameter desig-
nation on the PC
monitor
Parameter, measurement units
Possible values
Value after initial-
ization
1
2
3
4
Address
Address of the flow meter in the net-
work
1-247
1
Boudrate
Baud rate in the network of devices
1200, 2400, 4800
4800
Time and Date
Current device time and date
XX/XX/XX
(dd.mm.yy)
XX:XX:XX
(hh:mm:ss)
No changes
available
Switching mode
Clock transition between "Summer"
and "Winter" time
no switch
standard
user
standard
Data log clear
Reset of accumulated archives
...; start
...
Total flow reset
Zeroing of accumulated volume me-
ters
...; start
...
Table C.2. Service tab (Fig.10)
Median mean
Number of measurements to
determine the median value
1-21
7
Arithmetic mean
Number of measurements to
determine the arithmetic
mean
1-50
25
Inertion time, s
Minimum duration of USS
loss not fixed in the archives
0-255
10
Cutoff flowrate,
m
3
/h
Minimal flow rate cutoff
0.000-999.000
0.250
Fast reaction
Dynamic flow measurement
mode
On/Off
Off
Threshold, %Qmax
Criterion for quick installation
0-100
5
Contrast
Adjusting the indicator con-
trast
1-63
28
Updating time, s
Measured information update
period on the indicator
1-30
10
Worktime, s
Time after which the indicator
automatically turns off after
the last pressing of the button
1-255
30
Dimention
Flow rate dimension
m
3
/h; l/min
m
3
/h