24
Testing 2mm/3mm Jacketed Fibers with OFI 200
The following procedure may be used for testing 2mm/3mm jacketed fiber cables.
Note:
It is important to place the
plunger in the correct position. Please
refer to the illustration to the right.
1 Remove the plunger cover.
2 Make sure the plunger is oriented
as shown in the illustration. If
not, lift the plunger from the two
retaining pins.
3 Rotate the plunger such that
the wide groove ( labeled
“2mm/3mm”) is facing out and
will be used for aligning the
fiber.
4 Replace the plunger and plunger
cover.
5 Select the fiber to be tested.
6 Gently insert the fiber being tested into the fiber groove at the top of the OFI 200 head (refer
to Figure 2-1).
Plunger cover
Plunger
2mm
/3mm
250/900/RIB
OFI models plunger
top view, enlarged
Wide groove facing out
Retaining pins