AFHCAN 

www.afhcan.org 

Handy 

Guide 

Welch Allyn Otoscope 

How to Capture Images 

HG-0021 Welch Allyn Otoscope 

Copyright © 2015 Alaska Native Tribal Health Consortium. All rights reserved.  

Page 1 of 2 

 

The Welch Allyn Otoscope

 

1.  Fuse Drawer 

2.  Composite Video Outputs 

3.  S-Video Outputs 

4.  Fan & Grill 

5.  Power Output 

6.  Serial Number 

7.  Power Switch 

8.  Color Reset Button 

“White Balance” 

9.  Camera Port 

10. Light Port 

11. Light Control Knob 

12. Image Enhance Button 

13. Light Vent 

DO NOT BLOCK THE 

COOLING FAN

 

12 

10 

13 

11 

Operating the Welch Allyn Otoscope 

1. 

Turn on power.  When the power 
switch is on, the fan will power up and 
light will be emitted from the end of 
the Otoscope tip and through the light 
vent.

 

 

2. 

Ensure that the Light 
Control Knob is turned all 
the way to the right for 
maximum light output.

 

ONCE THE LAMP IS TURNED ON IT WILL TAKE AT LEAST 15 SECONDS TO FULLY LIGHT UP 

WHEN YOU TURN ON THE LAMP LEAVE IT ON FOR AT LEAST 60 SECONDS BEFORE TURNING OFF 

3. 

Perform a White Balance.  A white balance (color reset) should be performed once a week, after the lamp is 
replaced or whenever the image color appears distorted.  To obtain good clinical images the natural colors must 
be represented in the image.

 

To perform a White Balance: 

1.  Aim the Otoscope tip 1/2 to 1 inch away from a uniformly white sheet of paper. 

2.  While looking at the Otoscope light box, press the Color Reset button.  A green light will 

flash in the window to the right of the color reset button. 

3.  When it is done flashing, the white balance is complete.  Release the Color Reset button. 