Aethlabs microAeth MA200 Operating Manual Download Page 9

16 

 

microAeth

®

 MA Series MA200, MA300, MA350 

Operating Manual

microAeth

®

 MA Series MA200, MA300, MA350 

Operating Manual                  

  17

AethLabs

AethLabs

2.2.11. Shipping

The microAeth MA series instruments contain lithium-ion batteries and therefore must be packaged and 
shipped properly according to regulations for different shipping methods. Please refer to International 
Air Transport Association (IATA) regulations when shipping the microAeth as there are limits to how many 
microAeth can be shipped in a single box based on the battery size in each instrument. The microAeth 
MA200 contains a 3.6V, 3200mAh (11.52 Wh), 1 cell rechargeable lithium-ion battery. The microAeth MA300 
contains a 3.6V, 12800mAh (46.08 Wh), 4 cell rechargeable lithium-ion battery. The microAeth MA350 
contains a 3.6V, 12800mAh (46.08 Wh), 4 cell rechargeable lithium-ion battery. Please contact your shipping 
carrier for more information and for packing instructions.

2.3. Warranty

If product(s) were not purchased directly from AethLabs. please check with your reseller for Warranty details.

Disclaimer of Warranties; Limitation of Liability. Seller warrants that the Products sold hereunder, under normal 
use and service as described in the operator’s manual, shall be free from defects in workmanship and material 
for the lesser of (i) twelve (12) months, or (ii) the length of time specified in the operator’s manual, from the 
date of the shipment to the Buyer (“Warranty Period”). This Warranty Period is inclusive of any statutory 
warranty. Notwithstanding the foregoing, this limited warranty is subject to the following exclusions and 
exceptions: (i) air pumps are warranted only for ninety (90) days unless otherwise specified in the operator’s 
manuals; (ii) parts repaired or replaced as a result of repair services are warranted to be free from defects in 
workmanship and material, under normal use, for the later of (a) ninety (90) days from the date of shipment 
to the Buyer, or (b) the end of the Warranty Period; (iii) Seller does not provide any warranty on finished 
goods manufactured by others or on other consumable materials; and (iv) unless specifically authorized by 
separate writing by Seller, Seller makes no warranty with respect to, and shall have no liability in connection 
with, goods which are incorporated into other products or equipment, or which are modified by any person 
other than Seller. Seller agrees during the Warranty Period, to repair or replace, at Seller’s option, defective 
Products so as to cause the same to operate in substantial conformance with the published specifications 
thereof; provided that Buyer shall (i) promptly notify Seller in writing upon the discovery of any defect, which 
notice shall be provided during the Warranty Period and shall include the product model and serial number (if 
applicable) and details of the warranty claim, and (ii) prepay the shipment costs. Replacement parts may be 
new or refurbished, at the election of Seller. All replaced parts shall become the property of Seller. Shipment 
to customer of repaired or replacement Products shall be made in accordance with the delivery provisions set 
forth in these Terms. In no event shall Seller have any obligation to make repairs, replacements or corrections 
required, in whole or in part, as the result of: (i) normal wear and tear; (ii) accident, disaster or event of force 
majeure; (iii) misuse, fault or negligence of or by Buyer; (iv) use of the Products in a manner for which they 
were not designed; (v) causes external to the Products such as, but not limited to, water damage, impact 
damage from fall, power failure or electrical power surges, lack of maintenance; (vi) improper storage and 
handling of the Products; or (vii) use of the Products in combination with equipment or software not supplied 
by Seller. If Seller determines that Products for which Buyer has requested warranty services are not covered 
by the warranty hereunder, Buyer shall pay or reimburse Seller for all costs of investigating and responding 
to such request at Seller’s then prevailing time and materials rates. If Seller provides repair services or 
replacement parts that are not covered by this warranty, Buyer shall pay Seller therefore at Seller’s then 
prevailing time and materials rates. ANY INSTALLATION, MAINTENANCE, REPAIR, SERVICE, RELOCATION 
OR ALTERATION TO OR OF, OR OTHER TAMPERING WITH, THE PRODUCTS PERFORMED BY ANY 
PERSON OR ENTITY OTHER THAN SELLER WITHOUT SELLER S PRIOR WRITTEN APPROVAL, OR ANY 
USE OF REPLACEMENT PARTS NOT SUPPLIED BY SELLER, SHALL IMMEDIATELY VOID AND CANCEL 

ALL WARRANTIES WITH RESPECT TO THE AFFECTED PRODUCTS. THE OBLIGATIONS CREATED BY 
THIS WARRANTY STATEMENT TO REPAIR OR REPLACE A DEFECTIVE PRODUCT SHALL BE THE SOLE 
REMEDY OF BUYER IN THE EVENT OF A DEFECTIVE PRODUCT OR OTHERWISE FOR BREACH OF THIS 
WARRANTY OR THESE TERMS. EXCEPT AS EXPRESSLY PROVIDED IN THIS WARRANTY STATEMENT, 
SELLER DISCLAIMS ALL OTHER WARRANTIES, WHETHER EXPRESS OR IMPLIED, ORAL OR WRITTEN, 
WITH RESPECT TO THE PRODUCTS, INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES OF 
MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. SELLER DOES NOT WARRANT THAT 
THE PRODUCTS ARE ERROR-FREE OR WILL ACCOMPLISH ANY PARTICULAR RESULT. SELLER MAKES 
NO OTHER REPRESENTATION OR WARRANTY OF ANY KIND, EXPRESS OR IMPLIED, REGARDING THE 
PRODUCTS, INCLUDING, BUT NOT LIMITED TO, ANY WARRANTIES OF MERCHANTABILITY OR FITNESS 
FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL SELLER BE LIABLE FOR COSTS OF PROCUREMENT 
OF SUBSTITUTE GOODS AND SERVICES, LOSS OF USE OR LOST PROFITS, OR ANY OTHER SPECIAL, 
INCIDENTAL, INDIRECT OR CONSEQUENTIAL DAMAGES, ARISING OUT OF OR RELATED TO ANY SALE 
OF THE PRODUCTS OR THE USE OR PERFORMANCE OF THE PRODUCTS, HOWEVER CAUSED AND 
ON ANY THEORY OF LIABILITY, WHETHER IN AN ACTION FOR CONTRACT, TORT OR OTHERWISE, AND 
WHETHER OR NOT SELLER HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES.

2.4. End User License Agreement (EULA)

READ THIS END USER LICENSE AGREEMENT (“AGREEMENT”) CAREFULLY BEFORE USING THE 
ACCOMPANYING SOFTWARE AND ANY ASSOCIATED FIRMWARE (COLLECTIVELY, THE “SOFTWARE”). 
BY DOWNLOADING AND USING THE SOFTWARE, YOU ACKNOWLEDGE THAT YOU HAVE READ THIS 
AGREEMENT, THAT YOU UNDERSTAND IT, AND THAT YOU ACCEPT AND AGREE TO BE BOUND BY ITS 
TERMS. IF YOU ARE NOT WILLING TO BE BOUND BY THE TERMS OF THIS AGREEMENT, YOU SHOULD 
TERMINATE THE DOWNLOAD AND/OR INSTALLATION PROCESS, AND REFRAIN FROM ACCESSING OR 
USING THE SOFTWARE.

THE SOFTWARE IS COPYRIGHTED AND IS LICENSED TO YOU UNDER THIS AGREEMENT; IT IS NOT 
SOLD TO YOU. THIS AGREEMENT REPRESENTS THE ENTIRE AGREEMENT BETWEEN YOU AND 
MICROAETH CORPORATION DBA AETHLABS CONCERNING THE SOFTWARE, AND THIS AGREEMENT 
SUPERSEDES AND REPLACES ANY PRIOR PROPOSAL, REPRESENTATION, OR UNDERSTANDING YOU 
MAY HAVE HAD WITH AETHLABS RELATING TO THE SOFTWARE.

License Grant. microAeth Corporation DBA AethLabs (“Licensor”) hereby grants you a personal, non-
transferable and non-exclusive license to use (i) the Software, and (ii) any related documentation or electronic 
instructions accompanying the Software (the “Documentation”).

You obtain no right, title or ownership in any Software or Documentation, nor do you obtain any right to 
sublicense the Software. The Software may be used only as provided in either machine-readable object code 
form or machine-compressed form. This license is not a sale of the original or any copy of the Software. 
You may use the Documentation to assist in your use of the Software, but you may not make copies of the 
Documentation.
Use Restrictions. Except to the extent these restrictions are prohibited by applicable law or prohibited by 
the terms of any Open Source License, you must not, and must not allow any other person to: (a) use the 
Software to conduct comparative or competitive analyses, including benchmarking; (b) reverse engineer, 
decompile, reverse compile, reduce in human readable form or otherwise access the source code of the 
Software; (c) sublicense, rent, lease, modify, enhance, supplement, create derivative works from the Software; 

Summary of Contents for microAeth MA200

Page 1: ...microAeth MA Series MA200 MA300 MA350 Operating Manual 2018 AethLabs https aethlabs com Firmware v1 08 microAeth Manager v1 03 December 2018 rev03 ...

Page 2: ...3 5 Accessories 22 3 6 Replacement Parts 23 4 Measurements 24 4 1 Particles 24 4 1 1 DualSpot Loading Compensation beta 24 4 2 Temperature 24 4 3 Relative Humidity and Dewpoint 24 4 4 Altimeter Barometer 25 4 5 Accelerometer 25 4 6 GPS 25 4 7 Flow 25 5 Configuration and Operation 26 5 1 Overview 26 5 2 Recommendations for Best Use Practices 26 5 2 1 Instrument Settings Measurement Timebase and Flo...

Page 3: ...ta Download in the microAeth Manager 39 5 7 6 Exporting Data from the microAeth Manager local database 40 5 7 7 Delete All Data on the microAeth 41 5 8 Using the on board user interface 42 5 8 1 Overview 42 5 8 2 Configuration of Instrument Operating Parameters 42 5 8 2 1 Change Timebase 42 5 8 2 2 Change Flow 42 5 8 2 3 Change Tape Adv ATN 42 5 8 3 Operation and Status 43 5 8 3 1 Start Measuremen...

Page 4: ...ation Table Example 74 7 3 3 Test Flow Procedure 75 7 3 4 Flow Calibration Kit Setup 76 7 3 5 Flow Calibration Procedure 77 7 4 Optical Calibration Procedure 78 7 5 Installing Operating System Firmware 79 7 5 1 Upgrading firmware from version 1 05 to 1 08 79 7 5 2 Upgrading firmware from version 1 07 to 1 08 81 8 Technical Specifications 84 8 1 MA200 84 8 2 MA300 86 8 3 MA350 88 ...

Page 5: ...erosol The instruments have 5 analytical channels each operating at a different wavelength 880 nm 625 nm 528 nm 470 nm 375 nm Measurement at 880 nm is interpreted as concentration of Black Carbon BC Measurement at 375 nm is interpreted as Ultraviolet Particulate Matter UVPM indicative of organic sources such as woodsmoke tobacco and biomass burning The MA Series instruments have a number of import...

Page 6: ...ack port 8 USB mini B port 9 4 pin 3 3V TTL serial port 10 Serial number label 1 3 3 MA350 1 Enclosure door 2 Enclosure door padlock loop 3 Enclosure latches 2 4 Wall mounting kit bosses 4 5 Inlet port 6 Outlet port 7 Sealing circular connector for DC power in 3 3V TTL serial comms 8 Cap for sealing circular connector 9 Enclosure door gasket 10 Serial number label 11 USB mini B port 12 User interf...

Page 7: ...while the instrument is in use The lithium ion battery in the microAeth must be recycled or disposed of properly and separately from general or household waste in compliance with local laws and regulations The lithium ion battery contained in the microAeth and therefore the microAeth must be packaged and shipped properly Please see section 2 2 11 Shipping for more information Do not incinerate the...

Page 8: ...extreme temperatures Avoid exposing the microAeth connections and tubing to rapid and dramatic changes in temperature or humidity Care must be taken to identify installation and use conditions that might cause condensation of the sample aerosol stream or of the instrument itself Such condensation can cause instrument damage electrical shock or hazard 2 2 5 Consumables Filter tape cartridges will r...

Page 9: ...r Buyer shall pay or reimburse Seller for all costs of investigating and responding to such request at Seller s then prevailing time and materials rates If Seller provides repair services or replacement parts that are not covered by this warranty Buyer shall pay Seller therefore at Seller s then prevailing time and materials rates ANY INSTALLATION MAINTENANCE REPAIR SERVICE RELOCATION OR ALTERATIO...

Page 10: ... not do anything to jeopardize Licensor s or its licensors rights in the Licensor IP including to i copy modify merge or transmit Licensor IP ii register or attempt to register any competing intellectual property rights to the Licensor IP iii delete or tamper with any proprietary notices on or in the Licensor IP iv take or use any action that diminishes the value of any trademarks included in the ...

Page 11: ...bject matter hereof and supersedes all prior agreements oral or written with respect to the subject matter hereof 2 5 Regulatory and Compliance Notices This instrument meets the requirements of CE certification This device complies with Part 15 of the FCC Rules This Class A digital apparatus complies with Canadian ICES 003 Cet appareil numérique de la classe A est conforme à la norme NMB 003 du Ca...

Page 12: ...o bare leads cable This cable allows the microAeth MA Series instruments to send and receive 3 3v TTL serial data from the 4 pin serial port to and from other devices This cable is not included MA350 Sealed connector to bare leads cable This cable allows the microAeth MA350 to send and receive 3 3v TTL serial data to and from other devices This cable also allows the MA350 to be powered from an ext...

Page 13: ... in the column labeled Internal temp C The accuracy of the temperature measurement is typically 1 C at 25 C and typically 3 C over the temperature measurement range of 40 to 85 C 4 3 Relative Humidity and Dewpoint A relative humidity measurement of the sample air stream is made every Timebase This sensor is coupled to measurement sample spot sense2 In SingleSpot mode there is no active flow throug...

Page 14: ...of the instrument The following recommendations provide general guidelines 5 2 1 Instrument Settings Measurement Timebase and Flow Rate In order to get the best data from the microAeth for a sampling campaign we highly recommend that the instrument warm up for approximately 30 minutes so that it can equilibrate The microAeth can acquire data on six timebase settings 1 5 10 30 60 or 300 seconds The...

Page 15: ...or more targeted sampling can provide basic protection to the instrument inlet and prevents the possibility of light leakage into the instrument s sample chamber 5 3 2 Outlet Port The outlet port is a 10 32 UNF inch threaded port The outlet port and all connections to the outlet port must be properly protected from the environment There must be limited restriction to flow while protection from wat...

Page 16: ...chronized to the time on the computer Time synchronization will occur between the microAeth and the computer when the button is clicked in the microAeth Manager Time is ISO 8601 formatted It is very important to confirm the date and time of the microAeth before a sampling campaign 5 6 2 Date Time Timezone offset microAeth Manager only The timezone offset setting permits the user to select a timezo...

Page 17: ...d The tape advance ATN threshold setting permits the user to enter the attenuation ATN threshold natural number value of 1 to 100 that will trigger an automatic tape advance to a new filter sampling location during a sampling and measurement period The attenuation ATN threshold value will trigger a tape advance when the first of the wavelength measurements reaches this threshold NOTE The lowest wa...

Page 18: ...t takes to access the microAeth Manager local data base through the Manage data button will increase as the data base size increases 5 7 2 1 2 Notifications Section The notifications section displays notifications in the expandable and collapsible log Clicking the View all notifications button on the right of the main application window will toggle open the notifications log Clicking the same butt...

Page 19: ...in is highly recommended when using DualSpot loading compensation 5 7 3 7 Timebase The timebase dropdown can be used to select the measurement integrating time of 1 5 10 30 60 or 300 second s Select the desired value from the dropdown and click the checkmark to save and the x to cancel 5 7 3 8 Wavelengths The wavelengths dropdown can be used to select the measurement wavelength s used for particle...

Page 20: ...nloaded data sessions can then be exported to one csv file per session using the Data sessions in application section of the Manage data window Individual or multiple data sessions can be selected and exported When exporting long data sessions that span multiple calendar days the user can select to have the exported data session files split into multiple files based on the calendar days of that se...

Page 21: ...al microAeth Manager database on the computer 6 At the top of this window is the Data on device section to manage the data for this microAeth that is stored on the instrument At the bottom of this window is the Data sessions in application local database section where previously downloaded data sesions from the instrument that are stored in microAeth Manager local database are displayed 7 Select t...

Page 22: ...5 8 3 Operation and Status Before starting measurements it is recommended that the user verify all parameter settings A description of each operating parameter and its configuration is described in section 5 6 Instrument Operating Parameters Instrument settings can be viewed through the microAeth Manager software or through the serial terminal emulator interface using the Display All Settings menu...

Page 23: ...ton to select the Clamp Tape option The optical head of the instrument will move to its clamped position and stop resetting the optical head position Go to step 4 and try again to replace the filter tape cartridge 5 8 3 5 Calibrate Flow IMPORTANT Please read section 7 3 Flow Calibration and section 7 3 5 Flow Calibration Procedure before using this menu option The only way to perform a flow calibr...

Page 24: ... be displayed in the terminal emulator 13 Use the left and right buttons to scroll through the top level menu options to Display All Settings 14 Press the center button to select the Display All Settings option 15 Sent via Serial Port will display on the screen and all settings status and flow calibration table information will be displayed in the terminal emulator 5 8 3 12 Display FlowCal Info te...

Page 25: ...he Change Serial Baud menu or using the microAeth Manager and the Baud rate setting under Serial and Communications 5 9 1 AethLabs Serial Protocol The following AethLabs Serial Protocol commands can be used in polled serial data mode AethLabs Serial Protocol To issue a command send command carriage return CR ASCII code 0x0D Name Command Description Response Data request dr Requests most recent rec...

Page 26: ...II code 0x20 iii 3 Three digit instrument ID number SP 1 Space character ASCII code 0x20 000000 6 Spare field Set to all zeros SP 1 Space character ASCII code 0x20 ETX 1 End of text character ASCII code 0x03 BCC 2 Block checksum TOTAL 9 30 x A A is the number of gases AethLabs measurements being reported in the message Response Message for AethLabs Implementation of Bayern Hessen Protocol Complete...

Page 27: ...llected on the instrument in an exported file 6 Viewing and Analyzing Measurement Data 6 1 Data File Structure The data files exported through the microAeth Manager are plain text with the extension csv Each data file consists of a header containing descriptive information a line identifying the data columns and then a number of data lines with each item separated by a comma As the microAeth is st...

Page 28: ...r 100692 Measured pressure inside instrument case in pascals Internal temp C Degrees celsius Decimal number 35 375 Measured temperature inside instrument case in Celsius Exported Data File Structure Details Header Units Format Example Data Description Optical config DUAL or SINGLE SPOT XX XXXX XXXXX XXX XX DUALSPOT UV BLUE GREEN RED IR Identification of sampling mode SingleSpot or DualSpot and onl...

Page 29: ... BCc Calculated in DualSpot sampling mode UV BC1 Nanograms per cubic meter ng m3 Integer 23201 Calculated mass concentration for ultraviolet UV wavelength for sample spot sense1 in nanograms per cubic meter Calculated in SingleSpot and DualSpot sampling modes Exported Data File Structure Details Header Units Format Example Data Description UV BC2 Nanograms per cubic meter ng m3 Integer 24101 Calcu...

Page 30: ... Blue BC1 Blue BC2 Blue BCc Green BC1 Green BC2 Green BCc Red BC1 Red BC2 Red BCc IR BC1 IR BC2 IR BCc CKSUM UV IR Wavelengths Serial number Datum ID Session ID Data format version Firmware version Date Time GMT Timezone offset GPS lat GPS long GPS Speed Timebase Status Battery Accel X Accel Y Accel Z Tape position Flow setpoint Flow total Flow1 Flow2 Sample temp Sample RH Sample dewpoint Int pres...

Page 31: ...tenuation of ultraviolet UV wavelength for sample spot sense2 in attenuation units Calculated in DualSpot sampling mode Verbose Structure Details Header Units Format Example Data Description UV K Unitless Decimal number 0 0046992627903819 Calculated loading compensation factor for ultraviolet UV wavelength using sample spots sense1 and sense2 used to compensate BC1 value resulting in BCc Calculate...

Page 32: ...alculated in DualSpot sampling mode Blue BC1 Nanograms per cubic meter ng m3 Integer 24325 Calculated mass concentration for blue wavelength for sample spot sense1 in nanograms per cubic meter Calculated in SingleSpot and DualSpot sampling modes Blue BC2 Nanograms per cubic meter ng m3 Integer 24635 Calculated mass concentration for blue wavelength for sample spot sense2 in nanograms per cubic met...

Page 33: ...ration for ultraviolet UV wavelength for sample spot sense1 in nanograms per cubic meter Calculated in SingleSpot and DualSpot sampling modes UV BC2 Nanograms per cubic meter ng m3 Integer 24101 Calculated mass concentration for ultraviolet UV wavelength for sample spot sense2 in nanograms per cubic meter Calculated in DualSpot sampling mode IR BC1 Nanograms per cubic meter ng m3 Integer 23958 Cal...

Page 34: ...hLabs Serial Number Model and number MAxxx xxxx MA200 0011 Unique identification number for the Instrument Firmware Version Decimal number 1 08 Unique identification number for firmware running on instrument when measurement was recorded Name Alphanumeric MA Device User editable device name through microAeth Manager Timezone Offset Minutes Integer 480 Timezone offset from Coordinated Universal Tim...

Page 35: ...relative humidity of instrument sampling stream in percent Temperature Degrees celsius Decimal number 0 00 Measured temperature of instrument sampling stream in degrees celsius Acceleration X x xxx Y y yyy Z z zzz X 5 466 Y 18 466 Z 277 333 Measured instantaneous acceleration in the x axis y axis and z axis Satellites Whole number 0 Number of satellites that are visible to the instrument s interna...

Page 36: ...de TargetS SFlowT SFlow1 PumpS SSD TargetD DFlowT DFlow1 PumpD DSD S1 S2 0 0 00 124 99 124 03 0 0 00 0 00 124 90 123 90 0 0 01 0 00 1 43 93 180 30 179 20 69 2 04 45 75 181 79 164 51 49 2 46 1 41 2 58 82 197 69 197 76 88 1 65 58 74 197 48 175 92 65 1 82 1 42 3 74 38 215 58 216 97 112 1 21 74 90 215 49 190 49 87 1 48 1 41 4 89 79 231 66 235 80 140 0 94 90 07 231 28 203 94 111 1 21 1 41 5 104 83 245 ...

Page 37: ...sts A good complete flow calibration table is critical in order to obtain the best results from the microAeth To view the flow calibration table and the complete results from the previous flow calibration the on board user interface on the front of the microAeth is used to send all settings information of the instrument which includes the flow calibration table See section 5 8 3 11 Display All Set...

Page 38: ...9 48 348 23 680 0 20 1 38 17 0 00 0 00 0 00 0 0 00 283 32 357 21 359 09 789 0 14 1 37 18 0 00 0 00 0 00 0 0 00 0 00 0 00 0 00 0 0 00 0 00 19 0 00 0 00 0 00 0 0 00 0 00 0 00 0 00 0 0 00 0 00 20 0 00 0 00 0 00 0 0 00 0 00 0 00 0 00 0 0 00 0 00 21 0 00 0 00 0 00 0 0 00 0 00 0 00 0 00 0 0 00 0 00 FlowCal Results OK 35 35 7 3 3 Test Flow Procedure The on board user interface on the front of the microAe...

Page 39: ...mpling location The microAeth should be plugged into an external power source and its internal battery should be fully charged recommended battery remaining of at least 80 The only way to perform a flow calibration on the MA200 MA300 and MA350 is to use a microAeth MA Series Flow Calibration Kit provided by AethLabs 1 The microAeth must be at a new clean filter sampling location for the flow calib...

Page 40: ...microAeth can be used to upgrade the operating system firmware via the 4 pin serial port Upgrading the operating system firmware of the microAeth can be accomplished through a terminal emulator using the AethLabs 4 pin Serial to USB converter cable Before using the AethLabs 4 pin Serial to USB converter cable it may be required to install drivers from Future Technology Devices International Ltd FT...

Page 41: ...ff option 36 The memory on microAeth must be erased All data must be deleted Please follow the steps in section 5 7 7 Delete All Data on the microAeth to erase the memory 37 Set all settings using the microAeth Manager 7 5 2 Upgrading firmware from version 1 07 to 1 08 IMPORTANT READ FIRST data ON THE MICROAETH MAY NOT BE ACCESSIBLE AFTER UPGRADING FIRMWARE Before upgrading the firmware on the mic...

Page 42: ...ader will start 29 If the firmware transfer and upgrade are successful the data transfer will finish GOOD LOAD will be displayed self checks will complete SUCCESS and Shut Down will be displayed and the microAeth will shut down Once the screen backlight turns off the firmware upgrade is complete 30 The operating system firmware is upgraded on the microAeth but the flow calibration table should be ...

Page 43: ...ter spot s created on filter tape User selectable DualSpot or SingleSpot mode Environmental Sensors Accelerometer Relative Humidity Temperature Altimeter Barometer Dimensions L 136 75 mm 5 38 in W 85 mm 3 35 in D 35 75 mm 1 41 in Weight 420 grams 14 82 ounces Memory 16GB internal flash memory providing storage for 31 250 000 data lines 1 second timebase 361 days of data On board Interface Low Powe...

Page 44: ...d on filter tape User selectable DualSpot or SingleSpot mode Environmental Sensors Accelerometer Relative Humidity Temperature Altimeter Barometer Dimensions L 165 20 mm 6 50 in W 125 20 mm 4 93 in D 39 70 mm 1 56 in Weight 715 grams 25 22 ounces Memory 16GB internal flash memory providing storage for 31 250 000 data lines 1 second timebase 361 days of data On board Interface Low Power Screen 3 Bu...

Page 45: ...04 ounces Memory 16GB internal flash memory providing storage for 31 250 000 data lines 1 second timebase 361 days of data On board Interface Low Power Screen 3 Buttons Location services GPS with Internal Antenna Date Time Format ISO 8601 with satellite synchronization or manual computer synchronization Wireless 802 11 b g n Wi Fi with AES hardware encryption Bluetooth Low Energy Available in futu...

Reviews: