Auto Flow Calibration

1.  Always make sure that a filter strip is 

installed in the microAeth AE51 when 
it is operating.
 

2.  When exchanging a filter strip in the 

microAeth AE51, make sure that it is 
turned off.

 

3.  Install a clean, unused filter strip into 

the microAeth AE51. A pre-used filter 
with heavy loading may create an offset 
in the flow calibration table of the AE51. 
 
 
 
 
 
 
 
 
 

4.  Make sure that the flow calibration kit is 

connected to the AE51. 

5.  Connect the flowmeter and microAeth 

AE51 to the computer with the software 
installed. 

6.  Turn on the flowmeter and let it stabilize 

for at least 10 minutes before use. 

7.  Turn on the microAeth AE51. 

8.  Open the AE51 Auto Flow Calibration 

software.

 

 

 

 

 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

9.  Select the appropriate COM port from 

the dropdown menu in the External 
Flowmeter section of the software. 
There may be multiple COM port 
options depending on the computer 
and the peripherals connected to it. 

10. Click the Connect to Flowmeter button. 

If the port is already in use, a window 
will pop up to request that a different 
port is selected. If the status bar at the 
bottom of the software, does not show 
‘microAeth ON, Flow Meter Connected’ 
status, check your connections and 
ensure that communication with the 
microAeth AE51 and flowmeter has 
been initiated as previously described. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

11. Select the flow tolerance setting in the 

Auto Calibration section. 
 
The sofware will attempt to calibrate 
the internal flow system of the 
microAeth AE51 with 1% of the 
external flowmeter measurement. Some 
instruments, especially older units that 
have not had pump upgrade service 
by AethLabs, may need to use the 3% 
tolerance setting. Some instruments 
may have internal pumps that cannot 
reach the higher flow setpoints. 
Instruments not sold by AethLabs may 
require pump servicing by AethLabs 
in order to achieve flow rates of 200 
mlpm or higher. Contact AethLabs for 
additional information. 

12. Click the Calibrate button. 

 
The system will automatically cycle 
through all the setpoints of the flow 
calibration table of the microAeth 
AE51 (from 250 ml/min down to 0 ml/
min). The software will first check the 
maximum achievable flow rate of the 
instrument to determine what setpoints 
can be calibrated. If the internal pump 
cannot achieve the specified flow rate, 
the setpoint will not be calibrated. 
Contact AethLabs for a pump upgrade, 
replacement, or service. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

13. Click the View Instrument Values 

button. This will display all the values of 
the flow calibration table. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

14. Please check the values to make 

sure that as the flow setpoint 
increases from 0 to 250 ml/min, the 
pump drive and internal flowmeter 
values also increase.
 If this is not the 
case, please try again to calibrate the 
microAeth AE51 using the auto and 
manual flow calibrations. If this issue 
persists, please contact AethLabs for 
further assistance. 
 
 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 