microAeth
®
AE51 Flow Calibration Kit
Operating Manual
© 2016 AethLabs
https://aethlabs.com
Page 1: ...microAeth AE51 Flow Calibration Kit Operating Manual 2016 AethLabs https aethlabs com...
Page 2: ...ni DIN to 9 pin RS 232 with RS 232 to USB adapter 1 Flow Calibration Kit Documentation Software CD 1 TSI Flowmeter Operation and Service Manual 1 TSI RS 232 Serial Command Set Manual Flow Calibration...
Page 3: ...ad 7 Install a new filter strip by pressing and holding the release button and then inserting the new filter strip into the sample chamber opening 8 Make sure to push the new filter strip all the way...
Page 4: ...ter has been initiated as previously described 11 Select the flow tolerance setting in the Auto Calibration section The sofware will attempt to calibrate the internal flow system of the microAeth AE51...
Page 5: ...ated as previously described and disconnect the USB cable from the computer and reinsert it 9 Select the flow setpoint to calibrate from the dropdown menu in the Manual calibration section of the soft...
Page 6: ...try again to calibrate the AE51 If this issue persists please contact AethLabs for further assistance Save Instrument Values The Save Instrument Values button will prompt the user to select a location...