Medical electrical equipment needs special precautions regarding EMC for installation and use
according to the EMC information provided in this user manual.
Wireless communication equipment such as wireless home network devices, mobile phones,
cordless telephones and their base stations or walkie-talkies can affect this equipment and
should be kept at least a distance of 3.3m away from the equipment.
(Note: As indicated in Table 6 of IEC 60601-1-2:2007 for ME EQUIPMENT, a typical cell phone
with a maximum output power of 2 W yields d = 3,3 m at an IMMUNITY LEVEL of 3 V/m).
SYMBOLS
Contraindications
This describes situations where
you should not use the AEROSURE MEDIC™
LOT
-18°C
50°C
25%
85%
Warnings and cautions
Make sure you understand these
before using the AEROSURE MEDIC™
Type BF medical electrical equipment
Medical electrical equipment double insulated
Complies with European Medical
Devices Directive (93/42/EEC)
The Waste Electrical and Electronic Equipment Directive
(WEEE Directive). At the end of the product lifecycle, do not
throw this product into normal household waste, but take it
to a collection point for the recycling of electronic equipment
Legal manufacturer of the device
Consult instructions for use
Indicates the LOT number
Humidity and temperature limit
for storage and transportation
Indicates presence of NiMH battery.
(See page 47).
NiMH
Indicates the item model reference number
REF