MRK-D-0034 V2.0
Aeroqual AQM 65 User Guide
Page | 68
4.4.11 Add, remove or replace a gas module
Changing the module configuration can be done easily on site at the monitoring location. Module
replacement can be done with the AQM 65 power on.
A change in the module configuration will result in a change of inlet flow rate. It is important to
measure the inlet flow rate after the module change and document this in the instrument journal.
Tools required
Tube Cutter, Large Phillips head screw driver, AQM R7
Parts required
Tubing: CS Tube10 CS Tube11, power cable, communication cable,
2x luer cap
Frequency of
service
As required
Online
reference
training.aeroqual.com
: AQM 65 technical Training Section 5.2
Online video
Add a module
Remove a module
If a module is removed or added then a software change is required. The software module
configuration can be found in the configure instrument app. Sensor modules can be added or
removed using this tool, shown in Figure 4-21.
Figure 4-21 Change the module configuration in the software in the configure instrument app