4.4.3.3
Adjust the AQM 65 inlet flow rate and replacement of the gas sample pump.
If the AQM 65 inlet flow rate is found to be lower than expected, it could be because one of the
modules has a leak or is blocked.
To check this perform a leak check on the individual modules as described in Section 4.4.10.
If one of the modules is found to be blocked or has a leak then this will need to be corrected.
If all of the modules showed reduced flow, then it is likely that the gas flow pump is aging.
The flow adjustment valve can be used to provide more flow to the module inlets, and return the flow
rate to the expected flow rate.
Attach a flow meter to the inlet and begin measuring the inlet flow. Adjust pump flow adjustment valve
until the AQM 65 inlet flow rate has returned to the expected flowrate. The valve is shown in Figure
4-7.
Figure 4-7 The flow adaptor valve is used to adjust the AQM 65 inlet flow to the correct value.
If the correct flow rate cannot be set then it may be required to replace the gas sample pump. The
pump is part number
AQM R10.
The pump can easily be removed from the bracket for replacement. It
is important to use fresh Tygon Tube (CS Tube 10) when re-connecting the tube back on the pump.
This is because the barbs on the pump stretch the tube.
The online training website contains details on how to replace the gas sample pump.