4.4.2 Replace gas sample inlet filter
The gas inlet filter is held inside the enclosure at the top left. This must be changed to ensure
accurate readings, the filter is disposable and can be thrown away.
The filter should be changed when it becomes dirty as shown in Figure 4-4. The frequency of filer
changes depends upon local environmental conditions.
Tools required
None
Parts required
AQM R12 (bag of 25 filters)
Frequency of
service
Recommended
Monthly
(depends on local conditions)
See Section 4.6 for further discussion on filter change frequencies.
Online
reference
training.aeroqual.com
: AQM 65 technical Training Section 4.1
Online video
Do not turn off the AQM 65; this procedure can be done with the AQM 65 turned on.
NOTE:
It is very important to check the AQM 65 inlet flow rate is correct after changing the filter to
ensure there are no leaks.
Figure 4-4 The particle filter on the gas inlet can be easily replaced, it should be replaced when it gets
dirty as shown in the image on the right. (a new filter is shown on the left, dirty filter on the right)