Model 8310
IM-289A
Aeroflex / Weinschel
14
To configure the IEEE-488 bus address or serial communications parameters, select the appropriate switch setting
using the tables located in below (Figure 4).
GPIB SW1
Serial
Serial
Parameters
SP 1
SP Mode
Select
On = Serial parameters
Off = GPIB address
- - -
2
Echo
Echo Echo Enable
On = Echo received data
Off = No echo
- - -
3
HndshkSel
Handshaking Select
On = RTS/CTS
Off = XON/XOFF
A4 (16)
4
HndshkEna
Handshake Enable
On = Enabled
Off = Disabled
A3 (8)
5
ParitySel
Parity Select
On = Odd
Off = Even
A2 (4)
6
ParityEna
Parity Enable
On = Enabled
Off = Disabled
A1 (2)
A0 (1)
7
8
BR1
BR2
BaudRate Select (see below)
BaudRate Select
BR1 BR0
RATE
0 0
2400
0
1
9600
1
0
19200
1 1
38400
Note: The GPIB Bus address is selectable from 0 to 30 via the rear
panel dip switch. This switch is factory set to 10.
IEEE-488 Address Truth Table
Switch
Number
4 5 6 7 8
Decimal
Weight 16 8 4 2 1
Address:
0
1
2
3
4
5
6
7
8
9
10
20
30
0
0
0
0
0
0
0
0
0
0
0
1
1
0
0
0
0
0
0
0
0
1
1
1
0
1
0
0
0
0
1
1
1
1
0
0
0
1
1
0
0
1
1
0
0
1
1
0
0
1
0
1
0
1
0
1
0
1
0
1
0
1
0
0
0
Figure 4. Internal Dip Switch