Test Set Operation
Subject to Export Control, see Cover Page for details.
2 - 32
Fig. 2-39 Power vs. Time Tab view
2.4.4.E
P1 / P2 Measurements Window(s)
Rise Time(us) data field:
Displays the derived (Tri ggers: P1/P2 Rise Time: Start/Stop T rigger) Rise Time of the
captured DME pulse indicated.
Fall Time(us) data fi eld:
Displays the derived (T riggers: P1/P2 Fall Time: Start/Stop Trigger) Fall Time of the
captured DME pulse indicated.
Pu lse Wi dth(us) data fi eld:
Displays the derived (Tri ggers: Pulse Width/Spac ing: Width) Pulse W idth time of the
captured DME pulse indicated.
Po wer(dBm) data fi eld:
Displays the derived Pulse Power of the captured DME pulse indicated.
2.4.4.F
Measurements Window
PRF(Hz) data field:
Displays the derived Pulse Repetition Fr equency of the captur ed DME pulse pair.
Pu lse Spacing(us) data field:
Displays the derived (Tri ggers: Pulse Width/Spacing: S pacing) Pulse Spacing of the
captured DME pulse pair.
Percent Power d ata field:
Displays the derived (Inputs: Pct.Pwr Span) Percent Power (within the Occupied Band
Width) of the capt ured DME pulse pair.
Occupi ed Band Width(MHz) data field:
Displays the derived (Inputs: OBW Pct.) Occupied Band Wi dth of the captured DME
pulse pair.