![AEMC Simple Logger II L101 User Manual Download Page 17](http://html1.mh-extra.com/html/aemc/simple-logger-ii-l101/simple-logger-ii-l101_user-manual_2860728017.webp)
Simple Logger
®
II Series
15
MODEL
L642
ELECTRICAL
Channels
Two
Input
Two miniature thermocouple connectors
Measurement Type:
J
K
T
N
E
R
S
°F
°C
-346 to +2192
-210 to +1200
-328 to +2501
-200 to +1372
-418 to +752
-250 to +400
-328 to +2372
-200 to +1300
-238 to 1742
-150 to +950
32 to 3212
0 to 1767
32 to 3212
0 to 1767
Resolution
0.1
°
C/F < 1000
°
C/F; 1
°
≥
1000
°
C/F
Accuracy (J, K, T, N, E)
±(0.2% of R 0.6° C)
≤
(-100° C)
±(0.15% of R 0.6°C) > (-100° C),
≤
(+100°C)
±(0.1% of R 0.6°C) > (+100°C)
Accuracy (R, S)
±(0.15% of R 1.0°C) (0° to +100°C)
±(0.1% of R 1.0°C) > (+100°C)
Temperature
Coefficient:
± (0.02% of R 0.03°C)/ °C
or ± (0.02% of R 0.03°F) / °F between operating
temperature of -10°C to +18°C and +28°C to +50°C
Maximum Differential
Voltage
1V
(between T1 and T2)
Sample Rate
Maximum of 8 samples taken at storage interval
Storage Rate
Programmable from 5 sec to 1 day
Storage Modes
Start/Stop, FIFO and Extended Recording Mode* (XRM
™
)
Recording Length
15 minutes to 8 weeks, programmable using DataView
®
Memory
240,000 measurement (512KB). Recorded data is stored in non-
volatile memory and retained even if battery is low or removed.
Communication
USB 2.0 optically isolated
Power Source**
2x1.5V AA (LR6) alkaline batteries
Battery Life
100 hrs to >45 days (dependent on storage rate/recording length)
MECHANICAL
Dimensions
4.94 x 2.75 x 1.28" (125 x 70 x 32mm)
Weight (with battery)
7 oz (200g)
Case
UL94-V0
Vibration
IEC 68-2-6 (1.5mm, 10 to 55Hz)
Shock
IEC 68-2-27 (30G)
Drop
IEC 68-2-32 (1m)
ENVIRONMENTAL
Operating Temperature
14° to 122°F (-10° to 50°C)
Storage Temperature
-4° to 140°F (-20° to 60°C)
Relative Humidity
Up to 85% @ 95°F (35°C), Non-condensing
Altitude
2000m
SAFETY
Safety Rating
EN61010-1; 50V Cat. III; Pollution Degree 2