OPERATION 

Please make sure that you have already read and fully understand the WARNING section on page 1. 

Making Measurements with the AC Current Probe Model MN373 

  Connect the black lead of the  current probe to “common” and the red  lead to the AC v oltage input on   

your DMM or other v oltage-measuring instrument. The AC current probe has an output  of 10mV/A and 

1000mV/A. This means that with the probe in t he 10mV/A position, for 200A  AC in a conductor  around 

which the probe is clamped, 2V AC  will come out of the probe leads to  your DMM or instrument. Select 

the range which corresponds to the  measured current. If the current m agnitude is unknown, start with 

the probe in the 10mV/A pos ition and if the current is < 2 am

ps the probe can be  switched to  the 

1000mV/A position for the higher re solution. Clamp the probe around  the conductor. With the probe in 

the 10mV/A position take the r eading  on  the meter and multiply  it  by  100 to obtain the measured 

current (e.g, 160mV reading = 160 x 100 = 16,000mA or 16A). 

  For best accuracy, avoid if possible, the proximity of other conductors which may create noise. 

Tips for Making Precise Measurements 

  When using a current probe w ith a meter, it is im portant to select the r ange that prov ides the  best 

resolution. Failure to do this may result in measurement errors. 

  Make sure that probe jaw  mating surfaces are fr ee of dust and contamination.  Contaminants cause air 

gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for 

power measurement. 

MAINTENANCE: 

Warning

 

  For maintenance use only original factory replacement parts. 

  To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so. 

  To avoid electrical shock and/or damage to the  instrument, do not get w ater or other foreign agents 

into the probe 

Cleaning 

To ensure optimum performance, it is important to  keep the probe jaw mating surfaces clean at all times. 

Failure to do so may  result in error in readings.  To clean the probe jaw s, use very fine sand paper (fine 

600) to avoid scratching the jaw, then gently clean with a soft oiled cloth. 

REPAIR AND CALIBRATION 

You must contact our Serv ice Cent er for a Customer Serv ice Author ization number (CSA#). This  will 

ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# 

on the outside of the shipping container.  

Chauvin Arnoux

®

, Inc. d.b.a. AEMC

®

 Instruments 

 

15 Faraday Drive 

 Dover, NH 03820 USA 

(800) 945-2362  (Ext. 360) • (603) 749-6434  (Ext. 360) • [email protected] 

(Or contact your authorized distributor) 

NOTE: All customers must obtain a CSA# before returning any instrument. 

TECHNICAL AND SALES ASSISTANCE 

If y ou are ex periencing any  technical problems, or  require any  assistance w ith the proper use  or 

application of this instrument, please contact our technical hotline: 

(800) 343-1391 • (508) 698-2115 • [email protected] 

LIMITED WARRANTY 

The current probe is w arranted to  the owner for a period of one y ear from the dat e of original purchase 

against defects in manufacture. This  limited w arranty is giv en by  AEMC

®

 Instruments, not by  the 

distributor from whom it was purchased. This warranty is void if the unit has been tampered w ith, abused 

or if the defect is related to service not performed by AEMC

®

 Instruments.  

Full warranty coverage and product registration is available on our website at: 

www.aemc.com/warranty.html. 

Please print the online Warranty Coverage Information for your records. 

AC Current Probe 

Model MN373 

User Manual

 

 

DESCRIPTION 

The MN373 (Catalog #2116.28) is the la test in compact AC current pr obes. Designed to meet the most 

stringent demands in industry and electrical contracting, it also meets the latest safety and performance 

standards. The probe has a measur ement  range  up  to  240Arms, a range and resolution capable of 

displaying 10mV of output per amp of measured current, a voltmeter accuracy of 0.75% or better to take 

full advantage of the accuracy of the probe and an input impedance of 



1M

WARNING 

The safety warnings are prov ided to ensure t he safety of personnel and proper operation of the 

instrument. Read the instruction completely. 
•  Use caution on any circuit: potentially high voltages and currents may be present and may pose 

a shock hazard. 

•  Do not use the probe if damaged.  Always connect the current  probe to the  measuring device 

before it is connected around the conductor 

•  Do not use on non-insulated conductor w ith  a potential to ground greater than 600V C AT III 

pollution 2. Use extreme caution when clamping around bare conductors or bus bars. 

•  Before each use, inspect the probe; look for cracks in housing or output cable insulation. 
•  Do not use clamp in wet environment or in locations that hazardous gases exist. 
•  Do not use the probe anywhere beyond the tactile barrier. 

INTERNATIONAL ELECTRICAL SYMBOLS 

This symbol signifies that the current probe is protected by double or reinforced insulation. Use 

only factory specified replacement parts when servicing the instrument. 
This symbol signifies CAUTION! and requests that the user refer to the user manual before 

using the instrument. 
This is a type A current sensor. This symbol signifies that application around and removal 

from HAZARDOUS LIVE conductors is permitted. 

DEFINITION OF MEASUREMENT CATEGORIES 
CAT I:

  For measurements on circuits not  directly  connected to the AC  supply  w all outlet such as 

protected secondaries, signal level, and limited energy circuits. 

CAT II: 

For measurements performed on circuits directly  connected to the electrical  distribution sy stem. 

Examples are measurements on household appliances or portable tools. 

CAT III:

  For measurements performed in the building inst

allation at the distribution lev el such as on 

hardwired equipment in fixed installation and circuit breakers. 

CAT IV: 

For measurements performed at  the  primary  electrical supply  (<1000V) such as on primary  

overcurrent protection devices, ripple control units, or meters. 

RECEIVING YOUR SHIPMENT 

Upon receiving your shipment, make sure that the  contents are consis tent with the packing list. Notify  

your distributor of any missing items. If the equipment appears to be  damaged, file a claim immediately  

with the carrier and notify your distributor at once, giving a detailed description of any damage  

 

w

w

w

.

G

lo

ba

lTe

st

Su

pp

ly

.c

om

Fin

d Q

ua

lit

y P

ro

du

ct

s O

nli

ne

 a

t:

sa

le

s@

G

lo

ba

lTe

st

Su

pp

ly.c

om

Reviews: