MiniFlex
®
Flexible AC Current Probe
CHAPTER 3
SPECIFICATIONS
REFERENCE CONDITIONS
Quantity of influence
Reference values
Temperature
23 ± 5° C
Relative humidity
20 to 75% RH
Frequency of the signal measured
40 to 400Hz
Type of signal
Sinusoidal
External electric field
< 1V/m
External DC magnetic field (earth field)
< 40A/m
External AC magnetic field
None
Position of the conductor
Centered in the measurement coil
Shape of the measurement coil
Nearly circular
Input impedance of the display device connected to housing
≥
1M
W
MODEL
MF 300-6-2-10-HF
MF 300-10-2-10-HF
ELECTRICAL
Range
30/300A
Signal Output
100mV/10mV/A
Frequency Range
5Hz to 1MHz @ -3db
Frequency Limitation
none
Influence Of Conductor Positioning
1.5% typical, 3% max
Influence Of Conductor Positioning
In Sensor Against Handle
4% typical, 6% max
External Conductor Influence
35dB to 40dB on contact
Accuracy
± 1% +0.3A
Common Mode Rejection
100dB typical, 80dB min
Max peak factor
(1)
at I nominal
1.5
Residual noise at I = 0 (Arms)
(2)
0.3A
Max phase shift at 1kHz (°)
1
Max offset voltage (mV
DC
)
50
5
Max output voltage (Vpeak)
± 4.5
Output impedance (k
W
)
1
Power Source
9V alkaline battery (6LF22)
Battery Life
140 hrs continuous operation or 10,000 one minute measurements
Battery Indicator
When Green LED starts blinking, remaining life is approx 8 hours;
When LED is OFF, the battery needs to be replaced
(1): Peak factor PF = Vpeak/Vrms
(2): The residual noise affects the measurement uncertainty according to the formula:
(I measured x 0.01)
2
+ (residual noise)
2
I measured
global uncertainty =
(I measured ≠ 0)
If the current measured is zero, the uncertainty is equal to the residual noise.