16
Data Logger Model DL913 & DL914
3.2.3 Configuration
In addition to enabling and disabling Wi-Fi, Configuration mode
lets you
set three parameters (current range, aggregation period, and reset to default)
that affect how the instrument measures and displays data. If you cannot get
the configured router connection then you can reset to default which will reset
the Wi-Fi parameters. Then you can communicate using the default SSID in AP
mode with open authentication.
3.2.3.1 Current Range
The instrument provides two measurement ranges: 300 A and 3000 A.
The 300 A range provides better resolution but lower measurable current. You
can select the measurement range in Configuration mode.
1.
In Configuration mode
, use the Up ▲ or Down ▼ button to display the
Current Range (I RANGE) configuration screen.
This screen will also show 1, 2, or 3 turns to indicate the instrument’s
configuration for the number of sensor turns (wraps).
You can c
onfigure the instrument for the number of turns in the software
only.
2. Press
to toggle between 300 A and 3000 A.