Ground Resistance Tester Models 3705
- 11 -
Indicates the Auto-Off feature is deactivated and the instrument
stays on permanently (until turned off by the user). With no
symbol displayed, after 5 minutes of non-use the instrument will
automatically shut itself off. The Auto-Off feature is turned
On/Off by pressing Hold at power-up.
Display flashes this symbol (1/second) during a low battery
condition. Measurements are still possible. When this symbol is
displayed continuously, measurements are no longer possible.
Battery replacement is necessary.
Indicates the approximate percentage of useful battery life
remaining (0-100%). This function is displayed only upon initial
power-up when the ON push-button is held down for more than
2 seconds.
The overload symbol will illuminate when the measurement
reading is beyond the limits of the instrument:
Resistance > 1200
Ω
Current > 30.00A rms
The beeper is activated in current overload.