AEM 30-0300 Instruction Manual Download Page 10

10

7/12/2017 -  DOCUMENT NUMBER: 10-0300

© 2017 AEM Performance Electronics

30-0300 X-Series UEGO Gauge

A 2-1/6" (52mm) hole is required to mount the X-Series gauge.  A bracket and thumbscrews are provided to facilitate
installation into a panel or gauge pod.  In some cases, the gauge cup may be pushed into a mounting  hole  causing  an
interference fit strong enough to retain the gauge; the supplied rubber band may be fit to the gauge to create a tighter fit in
mounting holes slightly larger than 52mm.  It is, however, recommended that gauges be mounted securely  using  the
supplied bracket to ensure they never become loose and cause a hazard during vehicle operation.

Note: 

The gauge is not water-proof and should not be installed in a location with exposure to water or snow.  Damage

caused by water ingress will not be covered under warranty.

0-5V Analog Output

WHITE WIRE = Analog Po
BROWN WIRE = Analog Negative -

0-5V Analog Output Scaling Formulas

AFR

=

(2.3750 * Volts) + 7.3125

Lambda

=

(0.1621 * Volts) + 0.4990

The 0-5V analog output is suitable  for  output  to  devices
such as loggers or ECUs.  This differential output requires
special care to ensure proper  operation.    The  WHITE
signal wire should  be  connected  to  the  positive  of  the
analog input of the logging device  or  ECU;  the  BROWN
wire must be connected to the negative of the analog input
of the logging device or ECU.  If the logging device or ECU
does not have a differential analog input (both a dedicated
positive and negative terminal for the analog input) then
connect the BROWN wire to the shared signal ground.   If
the device does not have a dedicated signal ground then
as a last course of action, connect it to the power ground
of the logging device.

Important  Note

:  If  bench  testing  the  analog  output

outside of a vehicle, a multimeter's positive lead may be
connected to the WHITE wire however the BROWN  wire
must be  connected  to  BOTH  the  multimeter's  negative
lead AND power ground  going  to  the  X-Series  UEGO
device.  This connection is usually made by the circuitry
inside an ECU or data logger.

0-5V Analog Output Scaling Table

Volts

Lambda

AFR (Gasoline)

<0.50

SENSOR NOT READY

0.50

0.58

8.50

0.75

0.62

9.09

1.00

0.66

9.69

1.25

0.70

10.28

1.50

0.74

10.88

1.75

0.78

11.47

2.00

0.82

12.06

2.25

0.86

12.66

2.50

0.90

13.25

2.75

0.94

13.84

3.00

0.99

14.44

3.25

1.03

15.03

3.50

1.07

15.63

3.75

1.11

16.22

4.00

1.15

16.81

4.25

1.19

17.41

4.50

1.23

18.00

>4.50

SENSOR ERROR

AEMnet (CAN Bus) Output

WHITE WIRE WITH BLACK STRIPE =  / CANH
GREEN WIRE WITH BLACK STRIPE = AEMnet- / CANL

The AEMnet output is suitable for output to AEM devices such as the AQ-1 data logger or Infinity ECU.  The following
CAN configuration and message definition information is provided below to facilitate interface with third-party devices.

Message ID

If multiple gauges are connected to the same CAN bus then they must each be configured to have unique message IDs
using the gauge's menu system. Depress the MODE button to navigate to the CAN submenu, press the SEL button until
the desired ID is displayed then wait 3 seconds to accept setting; dOnE will appear to confirm your selection. The options
"ID 1", "ID 2", ..., "ID 16" correspond to the 29-bit identifiers 0x0000180,0x0000181, 0x0000180, ..., 0x000018F.

Bus Termination

All AEMnet/CAN networks must be terminated to have an
equivalent  of  approximately  60  Ohms  of  resistance. 
Generally, this means a 120 Ohm resistor connected in

Summary of Contents for 30-0300

Page 1: ...bus and serial data stream are also available for logging or mixture control purposes The X Series Digital Wideband technology is US Patent 9 575 030 Features X Digital Technology Free air or resistor...

Page 2: ...and a UEGO sensor harness B the harnesses are connected to the gauge as shown below The UEGO sensor harness simply needs to be routed to the UEGO sensor whereas the Power IO harness will require furth...

Page 3: ...ngine compartment Take care when routing sensor harness near hot exhaust components Apply strain reliefs and wire coverings as necessary Use a 5A inline fuse on the switched 12V power supply line Pin...

Page 4: ...e ability to perform a free air calibration is provided for users that wish to do so A free air calibration can be used as a sensor ages however new sensors should use the resistor calibration for bes...

Page 5: ...om for more information The faceplate may be reversible displaying alternative scalings on either side Reference the Operation section of this manual for details on how to switch the display mode when...

Page 6: ...e silver bezel may be replaced for the existing bezel at this time STEP 2 Separate the PCB 11 from the remaining components STEP 3 Slide the light guide 5 upward to remove it the buttons may fall out...

Page 7: ...slide the buttons 6 into the slots STEP 2 Stack the diffuser 4 faceplate 3 and lens 1 in order over the buttons and on to the light guide STEP 3 Reassemble the PCB and display stack with the bezel ma...

Page 8: ...unitless ratio that is fuel agnostic In other words a lambda reading of 1 0 is stochiometric for any fuel AFR Stoichiometric AFR Lambda In general an engine will have three areas of operation idle cru...

Page 9: ...ndle with caution UEGO sensors are very sensitive devices that must handled and installed with care Listed below are several contaminants and conditions which will shorten the life of the sensor or re...

Page 10: ...ted signal ground then as a last course of action connect it to the power ground of the logging device Important Note If bench testing the analog output outside of a vehicle a multimeter s positive le...

Page 11: ...erial RS 232 Output BLUE WIRE Serial Out The serial datastream is suitable for output to third party devices such as data loggers PCs or reflashed ECUs The X Series UEGO serial datastream is designed...

Page 12: ...the gasoline stoichiometric ratio of 14 65 which might be confusing for alternative fuels Remember that AFR lambda stoichiometric ratio For support contact AEM Technical Support at 1 800 423 0046 or...

Page 13: ...rating Temperature min 4 20 degF degC max 16V Supply 185 85 degF degC Sensor Temperature max sensor element 1706 930 degF degC Display Range min outer LED 0 55 lambda max outer LED 1 40 lambda min inn...

Reviews: