11. If an irregular heartbeat (IHB) from common arrhythmias is detected in the procedure
of blood pressure measurement, this sign
will be displayed. Under this condition,
the Electronic Sphygmomanometers can keep function, but the results may not be
accurate, it’s suggested that you consult with your physician for accurate assessment.
There are 2 conditions under which the signal of IHB will be displayed:
1) The coefficient of variation (CV) of the pulses is > 25 %.
2) The deviation of the following pulse period is ≥ 0.14 s, and the number of such
pulses amounts to more than 53 % of the total number of measured pulses.
12. Please do not use the cuff other than supplied by the manufacturer, otherwise it may
bring biocompatible hazard and might result in measurement error.
13. The monitor might not meet its performance specifications or cause a safety hazard
if stored or used outside the temperature and humidity ranges specified in the specifi-
cations.
14. Please do not share the cuff with other contagious persons to avoid cross-infection.
15. This equipment has been tested and found to comply with the limits for a Class B
digital device, pursuant to part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful interference
to radio communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful interference to
radio or television reception, which can be determined by turning the equipment off
and on, the user is encouraged to try to correct the interference by one or more of the
following measures:
- Reorient or relocate the receiving antenna.
- Increase the distance between the equipment and the receiver.
- Connect the device to an outlet on a circuit different from that to which the
receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
16. This blood pressure monitor is verified by auscultatory method. It is recommended
that you check annex B of ANSI/AAMI SP-10:2002+A1:2003+A2:2006 for details of
verification method if you need.
SETUP AND OPERATING PROCEDURES
1. BATTERY LOADING
a. Open battery cover at the back of the monitor.
b. Load four batteries of the type “AA”. Please pay attention to polarity.
c. Close the battery cover.