Sequence - Shock then No Shock Advised
(Fixed setting #1)
•
Press
on the remote
•
Press
F1
( )
•
On the LED display you should see the first bars as
green and the second, third and fourth bars as red
•
Press
on the remote. It is now set for Shock,
No Shock Advised.
To configure other shock scenarios, go to Page 19 in the
CR-T
Operating Instructions.
Set to Fully-Automatic
(no use of shock button) mode
•
Press
on the remote
•
Press
OP
( )
•
Use
to make the first bar in the LED display red
•
Press
•
Be sure to put plastic cover over shock button
on device
To configure for the semi-automatic setting, go to page 20
in the
CR-T
Operating Instructions.
For a faster demo, set CPR time to 15 seconds
•
Press
on the remote
•
Press
T1
( )
•
Move the red bar in the LED display to the first
position using the
button
•
Press
on the remote
To set other CPR time options, go to page 21 in the
CR-T
Operating Instructions.
Tip:
When using the remote, be sure it is within range of the
LED display/IRDA port so it will read commands.
Setting up your
CR-T
Slide the power switch on the right side of the device forward to turn on device.
OK symbol should now be in the readiness display.
Use the remote control to set your device into the following settings:
LIFEPAK
®
CR-T
AED TrAining SySTEm