background image

 

 

Copyright © 2005-2011 AEA Technology, Inc. 

5933 Sea Lion Place Suite 112, Carlsbad, CA 92010 

 Tel 800-258-7805 or +1(760) 931-8979 

 Fax +1(760) 931 8969 

www.aeatechnology.com

 

 

 

3.  Press ENTER to exit the Battery Menu and save the settings. 
 

Select popular settings and cable type  

 

These setting are user preferences from the F2 through F5 menus.  They include 
backlight timing, contrast, vertical scale (Ohms), horizontal scale (range), big 
numbers on/off, input channel, cable type, etc.  When completed go to the next 
steps to save all these settings. 
 

Save INSTRUMENT SETUP 

Steps 
1. Press F4 and select MEMORY ACTION. 
2. Select SAVE INST ONLY.  This will open a screen to enter a save name. 
3. Use the F keys and number keys to enter a name. 

4. Press ENTER to save the setup and enter a number 1-9 for the memory slot. 
5. Press ENTER to save the setup and exit. 
Only one instrument setup is required to recover the Battery Menu correctly.  
However, you can repeat this process to save up to 9 different setups, with 
different cable types and other setting as desired.  The saved settings can be 
penciled into the matrix in the Operator Manual in Appendix C. 

 

Soft Reset with RECALL INST 

Steps 
1.  If your 20/20 TDR is not performing correctly (i.e. Measurement screen has 
incomplete data, menu cursor will not move, etc.) perform the Soft Reset to restore 

the program operation. 
2.  Turn the TDR off. 
3.  Turn the TDR on and immediately press the ENTER key 4-5 times rapidly. 
4.  The TDR will perform the housekeeping necessary to restore operation and stop 
at a factory default Measurement Screen (normally 0-999 Ohms and Range about 

5120 ft). 
5.  Press the F4 key and select MEMORY ACTION 
6.  Select RECALL INST and enter a memory slot 1 through 9 to recall.  
7.  Press ENTER to recall the instrument setup. 
 

This action should avoid the issue of the battery menu changing the BATTERY TYPE 
to NONE or ALKALINE after a Soft Reset. Both those battery types will turn the 
TDR’s internal battery charger off.  As a last check you might wish to select the 
Battery Menu again to ensure the BATTERY TYPE is NICD NIMH.  Also, the EST. RUN 
TIME will display 0.0 HRS after a Soft Reset.  After 1-3 re-charging cycles the 
charger will learn the cell’s capacity and post valid EST. RUN TIME. 

  