R3267 Series Spectrum Analyzer Operation Manual (Vol.1)
1.5 System Checkout
1-11
Figure 1-8 Start-up Screen
6.
Attach the N-BNC adapter to the INPUT connector on the front panel and con-
nect the Input cable from the CAL OUT connector to the INPUT connector.
7.
Press SHIFT.
The SHIFT lamp lights.
8.
Press CONFIG(PRESET).
The default settings have now been reset.
The start-up screen is displayed as shown in Figure 1-8.
9.
Press CONFIG.
The Config menu is displayed.
Figure 1-9 Config Menu
10. Press more 1/2 and Selftest.
The Selftest menu is displayed.