POC-W211 Quick Start Manual

 

(3) Easy Setup 

Power On Computer: 

1) 

Make sure Equipotential Terminal is connected to 

equipotential ground in the hospital. 

2) 

Connect an 18Vdc, 100W adapter Power DC-IN. 

3) 

Press Power On/ Off. The button will turn green. 

 

Driver Installation: 

Advantech supports “one key” driver installation. A user can simply 

double click an icon to install all the drivers. 

 

1) 

Make sure the OS has been successfully installed. If you 

have difficulty installing an OS, please follow the instructions 

in the POC-W211 User Manual. 

2) 

Download the “one key” package from the Advantech Support 

website. Open the folder and double click “InstallAll.exe.” 

3) 

An Install dialog will appear. Follow the instructions shown in 

the dialog to finish the installation.   

 

* Please use a clean OS to install this auto installation; otherwise, 

unexpected errors may occur. 

* Should you only want to install specific drivers, please follow the 

instructions in the POC-W211 User Manual.   

 

Check Function Keys 

1) 

Press 

Volume Control

 to make sure sound volume can be 

adjusted.   

2) 

Press 

Touchscreen Function Control

 to make sure 

touchscreen functionality can be enabled/ disabled. If 

enabled, the button will turn green. 

3) 

Press 

Read Light Control

 to make sure the read lights can 

be turned on/ off. 

4) 

Press 

Brightness Control

 to make sure the brightness of the 

screen can be adjusted. 

(4) Optional Modules 

Optional modules are hidden in several places as shown. 

 

 

 

 

 

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

(5) Cleaning and Disinfecting 

During normal use, a POC (Point-of-Care) terminal may become 

dirty and should be cleaned regularly. 

 

1) 

Prepare a cleaning agent per manufacturer’s instruction or 

hospital protocol. 

2) 

Prepare a clean cloth  that has been moistened in a 

cleaning solution. 

3) 

Wipe the POC thoroughly with a clean wipe. 

 

Applicable Cleaning Agents List 

No 

Cleaning Agents 

No 

Cleaning Agents 

Cidex 

11 

Virkon and water 

(1:100) 

Isopropyl alcohol 

12 

DISPATCH 

Green tinctured 

soap 

13 

SDW 70C 

Windex 

14 

Trigene 

Alcohol 

15 

ANTIGONE WIPES 

Alcohol 70% 

16 

SporeClear 

Incidin plus 

17 

AntiGone 

Incidin liquid 

18 

MikroBac 

Mikrozid liquid 

 

 

10 

Chloride 1000ppm 

 

 

 

 

 

Left Side Door *

 

Right Side Door *

 

2 x USB 2.0 

Ports (Standard) 

Smart Card 

Reader 

RW-DVD ROM 

RFID Reader 

Bluetooth 

Module 

WiFi Module 

Battery Module 

* Do not immerse nor rinse a POC or its peripherals. 

* Do not spray cleaning agents on the chassis 

* Do not use disinfectants containing phenol 

*For model POC-W211-X01D-ATE only

 