Calibration Process

Before you can use PCLD-8811 to acquire data cor-

rectly, you must have offset and gain calibration.
Disconnect PCLD-8810 and open the DAQNavi to 

view the Device as Figure 1 .
1.  Select 

Device Setting

 and click 

Analog 

Calibration.

2.  Click 

Adjust Reference Voltage

 first and then 

click 

Auto Calibration Procedure

 to finish 

auto calibration.

Offset Calibration

Connect PCIE-1816H to the PCLD-8810,  and 

PCLD-8811 need to be installed onto which channel 

you want to calibrate AI 15 connect to AGND on 

PCLD-8810.

2.  Input a 4.999V to AI15 and Adjust VR4 until the 

value approach to 4.999. 

NOTE

: please adjust VR3 if the channel number is 

Even. (e.g. 0,2,4, ...)

Gain Calibration

After offset calibration finished, please execute gain 

calibration as follow:
1.  Click 

Device Setting

 and Setting 

Connect 

Type

 and 

Value Range Type

 of 

AI  Chan-

nel

1.  Click 

Instant AI

.

2.  Setting 

ChanStart

 =15 and 

ChanCount

 =1

3.  Click 

Start

 to continuously acquire data of AI 15.

4.  Adjust VR2 until the value approach to 0.000. 

NOTE:

 please adjust VR1 if the channel number is 

Even. (e.g. 0,2,4, ...)