Advantech HIT-W121 User Manual Download Page 28

Information on potential electromagnetic interference and advice on 

how to avoid or minimize such interference (IEC/EN 60601-1-2:2007) 

Guidance and manufacturer’s declaration – electromagnetic immunity 

The model HIT-W121 is intended for use in the electromagnetic environment specified below. The customer or the user of 
the model HIT-W121 should assure that it is used in such an environment. 

Immunity test 

IEC 60601 test level

Compliance 

level

Electromagnetic environment 

guidance

Conducted RF 
IEC 61000-4-6 

Radiated RF 
IEC 61000-4-3 

3 Vrms 
150 kHz to 80 MHz 

3 V/m 
80 MHz to 2,5 GHz 

3 Vrms 

3 V/m 

Portable and mobile RF communications 
equipment should be used no closer to any part 
of the model HIT-W121, including cables, than 
the recommended separation distance calculated 
from the equation applicable to the frequency of 
the transmitter. 

Recommended separation distance

                                                                                

d = 

1,2

P

                                                                                 

d = 

1,2

P

 80 MHz to 800 MHz 

                                                                                

d = 

2,3

P

 800 MHz to 2,5 GHz 

where 

P

is the maximum output power rating of 

the transmitter in watts (W) according to the 
transmitter manufacturer and 

d

is the 

recommended separation distance in metres (m). 

Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey, 

a

should be less than the compliance level in each 
frequency range. 

b

Interference may occur in the vicinity of 
equipment marked with the following symbol: 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption 
and reflection from structures, objects and people. 

a

  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 

radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To 
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be 
considered. If the measured field strength in the location in which the model HIT-W121 is used exceeds the applicable RF 
compliance level above, the model HIT-W121 should be observed to verify normal operation. If abnormal performance is 
observed, additional measures may be necessary, such as reorienting or relocating the model HIT-W121. 

b

  Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

FCC Caution: Any changes or modifications not expressly approved by the party responsible

for compliance could void the user's authority to operate this equipment.

FCC RF Radiation Exposure Statement: 

1.•This Transmitter must not be co-located or operating in conjunction with any other antenna or 
transmitter.
2.•This equipment complies with FCC RF radiation exposure limits set forth for an uncontrolled 
environment. This equipment should be installed and operated with a minimum distance of 20 
centimeters between the radiator and your body.

Summary of Contents for HIT-W121

Page 1: ...User Manual HIT W121 Computer...

Page 2: ...nterference in a residential installa tion This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful inte...

Page 3: ...se have the following information ready before you call Product name and serial number Description of your peripheral attachments Description of your software operating system version application soft...

Page 4: ...ied service personnel 14 If one of the following situations arises get the equipment checked by service personnel a The power cord or plug is damaged b Liquid has penetrated the equipment c The equipm...

Page 5: ...ndard IEC 60601 1 1 The unit is for exclusive interconnection with IEC 60601 1 certified equipment in the patient environment and IEC 60XXX certified equipment out side of the patient environment If i...

Page 6: ...HIT W121 User Manual vi...

Page 7: ...cations 2 2 1 Mechanical Specifications Terminal 2 2 2 Mechanical Drawing for Rear I O 2 3 External View 2 3 1 Front View 2 3 2 Side Bottom View 2 3 3 Rear View Chapter 3 Software Description 3 1 Wind...

Page 8: ...HIT W121 User Manual viii...

Page 9: ...1 Chapter 1 Introduction This chapter briefly introduces the HIT W121 product f Overview f System Configuration...

Page 10: ...e HIT W121 terminal also serves as an integrated hospital gateway device The HIT W121 equipped with an Intel Atom 1 6GHz Dual core pro cessor 11 6 W Full Flat touch screen onboard isolated Giga LAN po...

Page 11: ......

Page 12: ...HIT W121 User Manual 4...

Page 13: ...2 Chapter 2 Hardware Description This Chapter describes the hard ware features of the HIT W121 f General Specifications f Mechanical Specifications f External View...

Page 14: ...Operating Atmospheric Pressure 700 to 1060 hPa Storage Humidity 10 to 95 No Condensation Storage Atmospheric Pressure 700 to 1060 hPa Transportation Temperature 20 to 60 Transportation Humidity 10 to...

Page 15: ...ions 302 5 W x 220 25 H x 43 00 D mm f Carton Dimensions 500 W x 385 H x 160 f Mounting System VESA 75 75mm standard mounting hole f Net Weight 3 kg VESA Mount 75 75 mm or 100 100 mm SCREWS M4 10 mm C...

Page 16: ...2 2 2 Mechanical Drawing for Rear I O...

Page 17: ...2 3 External View 2 3 1 Front View Front View Terminal 2 3 2 Side Bottom View...

Page 18: ...2 3 3 Rear View Rear View Terminal 9...

Page 19: ......

Page 20: ...3 Chapter 3 Software Description This Chapter describes the soft ware features of the HIT W121 f Windows XP Embedded Software Specifications...

Page 21: ...LC880 Rev 5 10 0 5512 f LAN Marvell Yukon Ethernet Controller Rev 8 61 7 3 f Camera D Max UVC Webcam Rev 6 11 200 4 f Smartcard reader ALCOR micro USB smartcard reader Rev 5 11 9520 7 f Touch Elo Ttou...

Page 22: ...13 HIT W121 User Manual...

Page 23: ...HIT W121 User Manual 14...

Page 24: ...4 Chapter 4 Design Requirements This Chapter describes the design requirements of the HIT W121 f Environmental Specifications f Reliability...

Page 25: ...Less than 40 C 25 C ambient temperature front bezel Safety f CE EMI f FCC class B approved Vibration f 10 18 Hz 1 5 mm peak to peak displacement f 18 500 Hz 1 G acceleration 4 2 Reliability MTBF f 30...

Page 26: ...nded for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the model HIT W121 can help prevent electromagnetic interference by maintain...

Page 27: ...commercial or hospital environment Surge IEC 61000 4 5 r1 kV line s to line s r2 kV line s to earth r1 kV line s to line s r2 kV line s to earth Mains power quality should be that of a typical commerc...

Page 28: ...g symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection fr...