2.2
Installation Procedures
2.2.1
Connecting the Power Cord
The AVAS-402 can only be powered by connecting the power cord to its
power supply (FSP500M-80PA), and pressing the power button on the front
cover. Be sure to always handle the power cords by holding the plug ends
only.
Follow these procedures in order:
1.
Connect the female end of the power cord to the 3-pin AC male plug of
AVAS-402.
2.
Connect the 3-pin male plug of the power cord to an electrical outlet.
Figure 2.7 Connecting the power cord
2.2.2
Connecting the Ground Pin
Step 1. System ready and find the Equipotential Terminal on rear side of AVAS.
An Equipotential Terminal is provide to optionally connect to a hospital
ground/earth system
Figure 2-8: AVAS-402 Equipotential Terminal Pin
Step 2. Prepare the Grounding cable and the other terminal link to hospital
ground/earth system.