Advanced Instruments Inc.
Suitable fitting and 1/8” dia. 4-6 ft. in length of metal tubing to connect from the flow meter vent to tube fitting
designated SAMPLE IN.
Procedure:
1. With the span gas cylinder valve closed, install the regulator on the cylinder.
2. Open the regulator’s exit valve and partially open the pressure regulator’s control knob.
3. Open slightly the cylinder valve.
4. Loosen the nut connecting the regulator to the cylinder and bleed the pressure regulator.
5. Retighten the nut connecting the regulator to the cylinder
6. Adjust the regulator exit valve and slowly bleed the pressure regulator.
7. Open the cylinder valve completely.
8. Set the pressure between 5-30 psig using the pressure regulator’s control knob.
9. Caution: Do not exceed the recommended flow rate. Excessive flow rate could cause the backpressure on the
sensor
and may result in erroneous readings and permanent damage to the sensor.
18