Technical Specifications *
Accuracy:
< 2% of FS range under constant conditions
Analysis:
0-1%, 0-5%, 0-10%, 0-25% FS ranges; auto-ranging
or fixed single range
Application:
Oxygen analysis in inert, hydrocarbon, helium, hydro-
gen, mixed and acid (CO
2
) gas streams
Approvals:
Certified for use in hazardous areas - see lower right
UL: United States: UL 1203, UL 913, UL 508
Canada: CAN/CSA C22.2 No. 30-M1986,
CAN/CSA C22.2 No. 157-92,
CAN/CSA C22.2 No. 14-10
ATEX: Directive 94/9/EC
Area Classification: Certified for use in hazardous areas - see lower right
Calibration:
Max interval
—
3 months. Air calibrate with clean
source of certified span gas, compressed, or ambient
(20.9% O2) air on 0-25% range.
Compensation:
Temperature
Connections:
1/8" compression tube fittings
Controls:
Water resistant keypad; menu driven range selection,
calibration and system functions
Display:
Graphical LCD 2.75” x 1.375”; resolution 0.001%;
displays real time ambient temperature and pressure
Enclosure:
NEMA Type 3R for rain in outdoor applications (UL)
NEMA 4X (ATEX)
Flow:
Not flow sensitive; recommended flow rate 1-2 SCFH
Linearity:
±1% of full scale
Pressure:
Inlet - regulate to 5-30 psig to deliver 1-2 SCFH flow;
vent - atmospheric
Power:
18-24 VDC
Response Time:
90% of final reading in 10 seconds
Sample System:
Sample flow meter; options available, see other side
Sensitivity:
< 0.5% of FS range
Sensor Model:
GPR-11-32 for non-acid (CO2) gas streams
XLT-11-24 for gases containing > 0.5% CO2
Sensor Life:
GPR-11-32 32 months in air at 25°C and 1 atm
XLT-11-24 24 months in air at 25°C and 1 atm
Signal Output:
4-20mA
Operating Range: 5ºC to 45ºC (GPR sensor); -10
º
to 45
º
C (XLT)
Warranty:
12 months analyzer; 12 months sensor
Wetted Parts:
Stainless steel
Optional Equipment
Sample conditioning systems (see back page) - Contact factory
* Specifications subject to change without notice
Exia
UL Certified
File E343386
Class I, Division 1, Groups C and D
T4
T
amb
-20
⁰
C to +50
⁰
C
0080
ATEX Certified
-
Directive 94/9/EC
Examination Cert: INERIS 08ATEX0036
II 2 G
Ex d [ib] ib IIB T4
T
amb
-20
⁰
C to +50
⁰
C
UL or ATEX Certified for Hazardous Areas