MONROE ELECTROSTATIC FIELDMETER 282H

1. 

Turn the instrument ON.

2. 

Discharge your body by touching a grounded metal object.

3. 

Point the aperture toward a grounded object and press the ZERO button.

4. 

Aim the aperture toward the target surface at a distance of 1ft. Adjust the distance until the flashing beams of the 
LEDs inthe instrument converge. Read the voltage and polarity of the charged surface on the meter display. 

For additional measurements, repeat step four. 

Maintenance

The battery should be replaced annually, whenever you plan an extended period of unattended monitoring, or 
whenever "BAT" appears on the display for more than an instant. To obtain accurate and drift-free readings, the 
sensor plate and especially the area around the aperture must be kept absolutely clean at all times. Never touchthe 
aperture with anything—not even a cotton swab.