Adtech SCX 102 Instruction Manual Download Page 2

.  i 

6.0  sex 102 CALIBRATION 
The calibration can be performed in two methods. Section 6. 1 

describes the procedure when high precision input source, 5 

digit DVM and a 250 ohm± 0.01% shunt is available. 
Section  6.2  describes  an  alternate  method  of  differential 

measurement which is a bench test method and provides very 

high accuracy calibration. In this case the input and output are 

connected to provide differential error that must be nulled out. 

This procedure does not need high precision source or shunt. 

6.1  CALIBRATION WITH PRECISIONS COMPONENTS 
To perform a calibration check or recalibration of the instrument, 

follow the procedure below. 

A. Make sure the input output wiring is connected properly

per diagram on front page.

B.  The  input  signal  source  must  be  adjustable  from

0-100% in steps of at least 25%. The source must be

either precalibrated or a precision milliammeter should

be used to monitor the input.

C. Monitor the output across a 250n ±.01  % shunt
D. Set the input source to 4.000 mA and adjust the multi­

turn  control  marked  ZERO  to  provide  4.000 ±  mA

(1.000 vdc) within desired accuracy.

E. Set the input source to 20.000 ma and adjust the multi­

turn  potentiometer control marked SPAN  to provide

20.000 mA (5.000 vdc across 250 ohm).

F. Repeat steps D and E until desired accuracy is achieved.
G. Check linearity at input of 8.000, 12.000 and 16.000 mA.
H. This completes the calibration.

CAUBRATJON  CONNECTION 

WITH  PRECISION  COMPONENTS 

VOLTAGE 

METER 

6.2  CALIBRATION WITHOUT 

PRECISION COMPONENTS 

This method 

cannot 

be used when the isolator is connected in 

the actual circuits, because it needs connecting the input to the 

output which violates the 1/0 isolation. 

A. Remove  the  input/output  wiring  and  connect  the

instrument per figure on right below.

B. Set  the  input  source  to  approximately  4  made  and

adjust the multi-turn potentiometer marked ZERO  to 

provide 0.000 volt reading on the null meter.

C. Set  the  input source  to  approximately  20  made  and

adjust multi-turn potentiometer marked SPAN to provide

0.000 vdc on the null meter.

D. Repeat steps B and C until both readings provide 0.000

vdc on the null meter.

E. Check at inputs of 8, 12 and 16 ma for error if any.

7.0  TROUBLESHOOTING 
AD TECH does not recommend field trouble shooting of these 

device.  Make  sure  the  input  and  output  connections  are 

correctly  made  and recheck operation.  Return a defective 

instrument  to  ADTECH  factory  for  prompt  check  and 

recalibration. 

.. 

4.tt.  OM111 

(1S) 

VOLTAGE 

METER 

Note  1 :  Current  source  need  not  be  a  precision  source 

Note  2:  Voltage  at  Meter=%Error(of  20mA)  directly 

CALIBRATION  CONNECTIONS 

WITH  OUT  PRECISION  COMPONENTS 

-2-

190-A-SCX102-A

Reviews: