Adtech SCT 302 Instruction Manual Download Page 2

6.0  CONFIGURATION OF  INPUT AND OUTPUT 

All ADTE CH units are factory calibrated per P .0. instructions. 

Usually, a complete recalibration is not required unless it is 

required to change input types, output types or the range of 

the  unit.  Most  calibrations  will  only  require  a  ZERO  and 

SPAN adjustment. 

NOTE:  For  recalibration  to  the  existing  range  proceed  to 

section 6.3; for new input or output range proceed as follows. 

A. Open the case to gain access to pcb boards. The larger

pcb board is the input pcb and the smaller pcb is the

output/power pcb.

6.1  INPUT CONFIGURATION 

The  ADTECH  Model  SGT 302  Isolated  Voltage/Current 

Transmitter accepts both current and voltage inputs. 

The SGT 302 has been set at the factory per the input/output 

marked on the label. It is easy to change the type of input. All 

inputs listed in Table 1 can be changed by simply selecting 

the appropriate header jumpers J2 to JS on page 3. 

6.2  OUTPUT CONFIGURATION 

The output  has  been factory  set  as marked on the  serial 

number tag. However, it is easy to change the type of output 

if so  desired. All outputs  shown  in  Table  3  can  be  easily 

selected by selection of  jumpers J1 and J2 on the  output 

pc board. 

6.3  CALIBRATION 

To  perform  a  calibration  check  or  re-calibration  of  the 

instrument follow this procedure. 

A. Make sure the unit 1/0 wiring is properly connected and

that  the  correct  power  source  per  the  label  is  also

connected.  The instrument must be at normal power

for a minimum of 2 minutes before proceeding to B.

B. The input signal source(s) must be adjustable from Oto

100% in steps of 10% or at least 25%.  The source(s)

should  be either  precalibrated  or  an  accurate  meter

must be used to monitor the input(s).

C. The output may be monitored either as a direct voltage

for a voltage output signal or as a current that can be

read as a voltage across a resistor shunt e.g. 1-5 VDC

across 250 ohms fro 4-20 mA DC.

D. Set the input source to minimum input value and adjust

the multiturn potentiometer marked ZERO to provide

the minimum calibrated output (e.g.) 4.00 ma ± 0.01

ma de. Note: For zero based outputs it is better to set

input  at  10%  and  adjust  for  10%  output  for  ZERO

adjustment.

E. Set the input source to maximum value and adjust the

multiturn  potentiometer  marked  SPAN  to  provide

the maximum  calibrated  output  (e.g.)  20.00  ma

± 0.01 ma de.

F. Repeat  steps  D  and  E  until  readings  are  within

calibration.

G. The instrument should now be checked at 25-50-75%

of span minimum for linearity. 

H. This completes the calibration.

7.0  FIELD TROUBLE SHOOTING GUIDE 

This section offers a simple, first level trouble-shooting aid 

for an apparent instrument malfunction. 

SYMPTOM  CORRECTIVE  ACTION 

No output 

1. Check the input and output connections

carefully.

2. Check that the power supply polarity is

correct and that power is present on the

instrument terminals.

3. Check that the input source(s) is correct

and that it changes magnitude between

zero  and  full  scale  values  when  so

adjusted.

4. If the output is a current signal (4-20 ma,

etc.),  make  sure  the  output  loop  is

complete and that the correct meter range

is selected.

All external checks are complete.  Problem 

seems to be internal. 

The  instrument  is  made  of  small  components. 

Troubleshooting beyond the above may be difficult without 

special equipment. We do not recommend attempting repair 

of the unit in  the field. ADTECH offers a very responsive 

repair policy. Contact the ADTE CH factory for information on 

repair and return at 716-383-8280 or  716-383-8386 (FAX). 

-2-

190-A-000076- B

Reviews: