3.2 Warning Information
•
You should read and understand the contents of the instructions carefully
before operating the equipment.
•
Operation should follow all warnings and cautions on the instrument .
•
The power socket cable should be in accordance with the standard
configuration and the ground wire must be secure .
•
Delivery head shall not be stacked with too heavy items (max load 2.5kg) .
•
Air pressure of the system has been set before leaving factory, non-
professionals are not free to adjust.
•
Turn off the main switch of water, gas, power when get off work every day.
•
The handpiece shall supply water only when the foot switch is in the state
of water supply. (only for 4-pin handpiece tube)
•
Hospitals or individual clinics who use the equipment must be equipped
with or connected to an amalgam separation device in the waste discharge
system .
•
The water and air must be exhausted and then cut off the water source and
electricity when maintaning or repairing; switch on the power, water and
air before the use.
•
The 3-way syringe handle and syringe tip should be put into the disinfection
bag to disinfect, and then 134
℃
(pressure 29.8psi) high pressure steam
disinfection, disinfection time is no less than 3 minutes.
•
Please read the user’s guide before cleaning and disinfecting the high speed
turbine and pneumatic micromotor.
•
The cover clean with wet cloth and neutral detergent to ensure smooth and
elastic leather surface free from corrosion; ABS parts clean with a soft cloth
dampened with soap water; the metal baking parts wash with soft cloth
dipped in soapy water or wipe with a wax spray.
•
Take an extra care with a fiber-optic handpiece to avoid damaging the
shining tip and make sure the tip will not come into contact with the treated
mixture, keep a certain distance or use a transparent sheet before the first 5
seconds of treatment.
•
Any traces left by the mixture on the tool must be removed immediately,
remove the fiber-optic handpiece and clean it with a cloth dipped in alcohol.
•
Don’t point the light of the fiber-optic handpiece to the patient's eyes for the
light may harm to patient's eyes,even cause temporary blindness!