5.2 Fix the chair 

Attentions: 

1.The chair must be fixed on the floor for ensuring  stability, for it can 
be seriously damaged  by wrong fixation. 
2.Check out the material of the floor  on the installing place. Please 
contact the authorized contractor to thickening the floor if its 
thickness is less than 82 mm. 

1

Power on and activate the unit. 

2

Step the button            on foot control , raise the unit. 

3

Unit switched off and power disconnected. 

 4

Remove the cover and mark the installing position , then move the unit away. 

Drill on the marked position, press the explosive bolt, then return the unit to its 

position and tighten the screw. 

Power switch 

Installation manual of AJ16US dental chair 

Summary of Contents for AJ16US

Page 1: ...P N 8021442 Version B Date 2017 08 01 Installation manual of AJ16US dental chair http www adsequip com ...

Page 2: ...g ADS does not warrant that this document is error free No part of this document may be copied reproduced altered or transmitted in any form or by any means electronic or mechanical including photocopying recording or by any information storage and retrieval system without prior written permission from ADS The document is intended to be used in conjunction with the equipment Comments and feedback ...

Page 3: ...5 4 Install the headrest 10 5 5 Install backrest upholstery 10 5 6 Install seat upholstery 11 6 Operations 12 6 1 Power 12 6 2 Chair positioning 13 6 2 1 Operate by foot control 13 6 2 2 Control panel and function set 14 6 2 3 System state 14 6 2 4 Initial settings of the chair 15 6 2 5 Initial settings description 16 6 2 6 Chair positions description 17 6 3 Dual articulating headrest 18 6 4 Armre...

Page 4: ... 1 Introduction 1 1 Product information 1 2 Symbols explanation Table 1 Installation manual of AJ16US dental chair Name Dental chair Model AJ16US Working conditions Ambient temperature 5 40 Relative humidity 80 Air pressure 12 5 15 4psi Nominal voltage AC110 V 22 V Nominal frequency 60 Hz 1 Hz P N 8021442 ...

Page 5: ...ent Notice Avoid electric shock Backrest Down Ground protection Backrest Up Original position Chair Down Set Chair Up Spitting position Program Chair Position 1 Program Chair Position 2 Program Chair Position 3 P 0 Installation manual of AJ16US dental chair ...

Page 6: ...good balance for usability simplicity and comfortability Figure 1 AJ16 US Dental Chair Item Component Item Component 1 Headrest 6 Armrest button 2 Backrest 7 Touch pad control panel 3 Armrest 8 Lifting arm 4 Seat 9 Junction box 5 Foot control 8 9 6 5 7 1 2 4 3 Installation manual of AJ16US dental chair ...

Page 7: ... P N 8021442 3 Paremeters 3 1 Specifications Load bearing Patient bearing Max 135kg Tip Specifications are subject to change without notice 3 2 Size unit inches Installation manual of AJ16US dental chair ...

Page 8: ...vice should not operate in the environment of high temperature high humidity vibration flammability unbalance or dusty Only the authorized spare parts can be used in this device This dental unit should not come into contact with any object in its operational range of movement This device can not be used in the environment of strong magnetic electromagnetic Leather maintenance 1 No organic solvent ...

Page 9: ...al drill Pincers Hammer Scissors Leveling instrument Flat bladed Screw driver Phillip screw driver Allen wrench Spanner Hexagon socket wrench The tools and instruments need to prepare before installation Installation manual of AJ16US dental chair ...

Page 10: ...ir Holds Installation manual of AJ16US dental chair 1 Remove all debris and cardboards around the chair 2 Hold the handrail and the front of the frame uplift the chair and place the chair on the right place 3 Remove binding belt and packages 4 Remove the front shell Size inches ...

Page 11: ...zed contractor to thickening the floor if its thickness is less than 82 mm 1 Power on and activate the unit 2 Step the button on foot control raise the unit 3 Unit switched off and power disconnected 4 Remove the cover and mark the installing position then move the unit away Drill on the marked position press the explosive bolt then return the unit to its position and tighten the screw Power switc...

Page 12: ...n socket cap screw assembly 4 GFM 1214 10 Slide bearing 5 Push rod 1 Figure 2 P N 8021442 2 3 5 4 Pass the fixing shaft through the backrest push rod fix the shaft on backrest with the cap screw assembly Tool 5mm Allen wrench Installation manual of AJ16US dental chair ...

Page 13: ... into the slot of the backrest Figure 3 5 5 Install backrest upholstery 1 Metal backrest 2 Fixing shaft 3 Backrest upholstery When installing the upholstery you should put the upholstery on the backrest with the shaft head pass through corresponding keyholes of the upholstery then push the upholstery in force till it s fixed ...

Page 14: ...lstery onto the chair frame lock the bracket and the rear end of the chair frame together and tighten the two screws on the front end of the chair frame Figure 5 P N 8021442 Installation manual of AJ16US dental chair C A B A Fixing bracket B Chair frame C M5 16 hex cap screws ...

Page 15: ...he power box controls power supplying of the dental chair and its delivery system Switch on and you can get the power switch off for not using in a long term or before maintenance 12 P N 8021442 Installation manual of AJ16US dental chair Figure 6 ...

Page 16: ... on the foot control till the chair is in best position Move the backrest by pressing backrest up down button step on the left right arrow till the chair is in best position 13 Figure 7 P N 8021442 Original position Program Chair Position 1 Program Chair Position 2 Set Spitting position Backrest down Backrest up Chair up Chair down 0 1 2 LP P Installation manual of AJ16US dental chair ...

Page 17: ...position Backrest down Backrest up Chair up Chair down 6 2 2 Control panel and function set 14 Figure 8 P N 8021442 6 2 3 System state If the indicator light is bright on the control panel it indicating that the system is normal and you can operate normally 0 LP P Installation manual of AJ16US dental chair ...

Page 18: ...he higher limit switch is triggered System initialize automatically it can be normally performed after initialization finished Hydraulic system initial settings specification 1 Press P button about 12 second until the buzzer sound 2 short beeps enter original setup status press P1 the buzzer sound 1short beep escape setup status This steep eliminates the original stroke parameters 2 Press P button...

Page 19: ...ters the setup procedure Press the P1 button and then press the P button one time again beep1 2times setup finished and escape the setup procedure P2 P3 position setup are same step as the P1 P P Installation manual of AJ16US dental chair Caution Hydraulic system shall have the initial settings in following situations First installation or replacement of mainboard Insert replace repair any positio...

Page 20: ...tal chair a Backrest Up Press this button continuously and the backrest will move upwards When the backrest reach the desired position stop pressing the button and the backrest will stop immediately b Backrest Down Press this button continuously and the backrest will move downwards When the backrest reach the desired position stop pressing the button and the backrest will stop immediately NOTE If ...

Page 21: ...AJ16US dental chair e Original position Press the button the chair and backrest will move to the original position automatically Original position backrest highest the chair lowest and light off Press this button again the chair returns to former position and light on The chair shall stop at a new position by halfway stop 6 3 Dual articulating headrest Figure 8 Headrest adjustment A B C A Headrest...

Page 22: ...eposition the headrest raise or descend the chair to adjust heights Patient s position should also be adjusted to put the wheelchair and dental chair back to back together 3 Make sure the wheelchair is locked 6 4 Armrest The armrest is easy to operate which provides convenience to users Press the button you can rotate the armrest forward or backward Button Figure10 Armrest positioning 19 ...

Page 23: ... to tighten the brake when the chair gets to desired angle You should adjust the tension of handle if the chair can not be fixed or the handle is hard to move 1 Loosen the set screw by an Allen wrench 2 Rotate the handle to adjust tension clockwise to tighten counterclockwise to loosen 3 Tighten the screw after adjustment Figure 13 Handle adjustment Installation manual of AJ16US dental chair ...

Page 24: ...holstery To take off the upholstery from backrest you should hold the bottom edge of gasket firmly and lift it up 1 inch or 25 mm then lift the upholstery from the backrest of the chair Figure14 To reinstall the upholstery you can put the keyhole of gasket on nail head then push till the gasket is fixed Figure 14 Remove the upholstery P N 8021442 Installation manual of AJ16US dental chair ...

Page 25: ...holstery Loose the screw at the front end of the upholstery push the upholstery forward then remove the seat upholstery Figure 15 Remove the seat upholstery P N 8021442 Installation manual of AJ16US dental chair ...

Page 26: ...een reserved for dentist unit 9 1 4 black reserved for dentist unit 10 1 4 blue reserved for dentist unit 11 1 4 red reserved for dentist unit 12 Control wire for delivery head 13 7 core power cord for delivery head 14 24V power cord for dental light 15 Ground wire for delivery head 16 1 4 green reservd for assistant s control 17 1 4 yellow reservd for assistant s control 18 Control wire for assis...

Page 27: ...ct to delivery system 7 1 8 yellow connect to delivery system 8 1 8 green connect to delivery system 9 1 4 black connect to delivery system 10 1 4 blue connect to delivery system 11 1 4 red connect to delivery system 12 Control wire for delivery head 13 7 core power cord for delivery head 14 24V power cord for dental light 15 Ground wire for delivery head 16 1 4 blue connect to side box 17 1 4 gre...

Page 28: ...ery head 8 1 4 Black for delivery head 9 Control wire for delivery head 10 24V 7 core power cord for delivery head 11 Ground wire for delivery head 12 Control wire for assistant control 13 24V power cord for dental light 14 1 8 Greeen for side box 15 1 4 Yellow for side box 16 1 4 Blue for side box 17 Connect cuspidor Steel winding tube 18 Connect side box filter Screw thread tube Lines of Beyond3...

Page 29: ...26 P N 8021442 12 Control wire for delivery head 15 Ground wire for delivery head 14 24V power cord for dental light Installation manual of AJ16US dental chair Lines of Classic 100 101 ...

Page 30: ...on The dental system suggests disinfecting the leather once a month with a 3 hydrogen peroxide solution leather cleaner or 3 muti enzyme cleaner based on its instruction to ensure effective disinfection and then fully wipe it with water Caution Do not disinfect the upholstery with alcohol potentiated glutaraldehyde disinfectant liquid bleach Benzammonium bromide Benzalkonium chloride disinfectant ...

Page 31: ...28 P N 8021442 Connecting diagram Hydraulic system Installation manual of AJ16US dental chair 10 Connecting diagrams ...

Page 32: ...e the right to make any alterations which may be due to technical improvements ADS Dental System Inc Add 1590 S Milliken Ave unit A Ontario CA 91761 Tel 626 620 0456 E mail ads adsequip com http www adsequip com ...

Reviews: