User Manual arsos
®
filius
+ dexos
light wds
Item No. 111 510-DL-wds
Ulcer Pressure Relief System
Assembly and Operation instructions
Health. Security. Independence.
Mattress arsos® filius
Power unit
dexos® light wds
Page 1: ...r Manual arsos filius dexos light wds Item No 111 510 DL wds Ulcer Pressure Relief System Assembly and Operation instructions Health Security Independence Mattress arsos filius Power unit dexos light...
Page 2: ...page 5 4 Explanation of indications and functions page 5 5 Installation arsos filius page 6 6 Installation dexos light wds page 6 7 Other settings page 10 8 Auxiliary functions page 10 9 Cleaning ins...
Page 3: ...ttress can be converted into a dynamic system through connecting the mattress to different control power units e g dexos light wds or dexos autoplus The composition of the system can be considered a s...
Page 4: ...XPLOSION DANGER Warning Note The compressor is under power as soon as the power supply plug is put in Before opening the housing of the compressor the power plug is to be unplugged Repairs must be per...
Page 5: ...results in an as large as possible surface and maximum pressure relaxation The reinforcements of the foam edges are an advantage for the sense of feeling and the mobility of the patient Touching the...
Page 6: ...tely The dual stretch and breathing cover is to be fixed to the bed base with the zipper on both sides Before putting the patient on the mattress you can fill air into the system with a hand pump As t...
Page 7: ...um and maximum level The unit starts up in alternating mode with a 5 minute cycle The red alarm light 5 is flashing and the system will start to be filled with air The red alarm light is put out of ci...
Page 8: ...button 2 lower When pressure is too low hammock effect press button 2 higher Please note that status of the air cells is to be checked daily and whenever required adjustment is needed Each time when...
Page 9: ...xtended pressure in the sacral area Optimal support of the patient Knee joints lumbar and cranial lordosis are supported Principle of alternating pressure every second cells pair deflated showing grou...
Page 10: ...change from green to blue alternate In alternate mode 2 out of 3 cells are filled with air to the set pressure level One cell remains without pressure Don t forget to increase the pressure when switch...
Page 11: ...ry function The unit has a memory function This means when the power is interrupted the unit will remember the set values up to about 24 hours Thus after transfer of a patient connecting to the power...
Page 12: ...mperature of 75 C should be kept In case of an arsos filius Mattress with an alternating power unit this should be cleaned with an damp cloth and soap detergent or a mild non aldehyds cleaning deterge...
Page 13: ...operate Power supply plug plug and partition blue lamp at defect power cord socket Check the the switch is not defect safety fuse safety fuse illuminated defect Red alarm light Defect air cell Check...
Page 14: ...cording to Braden according to EPUAP Patients with Patients with 10 kg 50 body weight and 50 kg body weight Static Pressure ulcers prevention and Instable fractures soft bedding therapy support up los...
Page 15: ...shable adequate detergents dryer can be used to 50 C Warranty only valid in case of production or material failures 12 Technical data 13 air foam combinations Item No Description dexos light dexos aut...
Page 16: ...s Phone 31 0 13 5772 088 Fax 31 0 13 5772 648 E mail info adlmedical nl Internet www adlmedical nl ADL GmbH Lise Meitner Stra e 3 48161 Muenster Germany Phone 49 0 2534 6224 0 Fax 49 0 2534 64270 ADL...