Adixen ASM 142 Operating Manual Download Page 10

10

 

Internal

:  The internal autocalibration is automatically activated  

 

during the start-up process. It doesn’t require any  

 

operator action. Thanks to the initial autocalibration,  

 

the leak detector can be immediately operational.

Internal autocalibration on request: it can be started by the 

operator whenever needed (the unit has to be off-cycle).

The result of the autocalibration process is displayed.

 

External:

  The external autocalibration allows direct readout in cases of operation with  

 

an auxiliary pumping system.

F1

F3

F2

F4

READY FOR CYCLE

INLET VENT  : off

F1

F3

F2

F4

READY FOR CYCLE

INLET VENT  : off

F1

F3

F2

F4

AUTOCAL IN PROGRESS
ELECTR. ZERO ADJUST.

80

To perform an external calibration 

 

C 303

AUTOCAL

Calibration

Summary of Contents for ASM 142

Page 1: ... GRAPH D Detector connections ASM 142 142 D Operator interface ASM GRAPH GRAPH D Operator interface Start up User interface level Test cycles Helium signal analog scale display Audio alarm Zero function References refer to a specific chapter of the Operating manual For further information please refer to Operating manual supplied with your unit 9 10 11 Condended Manual ...

Page 2: ...ft side Right side Printer interface headphone loudspeaker option RS232 interface option I O interface option Quick connector exhaust Long distance sniffer quick disconnect Neutral gas purge ASM 142 D and ASM GRAPH D only Detector connections ...

Page 3: ...lay 12 Helium signal analogic scale ON OFF indicator 13 Helium signal Zero scale ON OFF indicator 14 Correction factor COR indicator applied to digital display 15 Units ON OFF indicator 16 Helium signal digital display 17 Alphanumeric display 4 lines x 20 characters 18 Parameter function keys 1 key per display line 19 Modification access keys 4 keys 20 NEXT next display parameter circular function...

Page 4: ... parameter circular function 21 22 Plus or minus value adjustment parameter selection audio volume adjustment keys 23 NEXT of previously displayed values cancels tempo rary inputs 24 Menu selection access key 4 keys 25 SET POINT menu selection key 26 SPECTRO calibration and analyzer cell configuration menu selection key 27 MAINTENANCE menu selection key 28 OTHER menus selection key test mode selec...

Page 5: ...he unit please fill up the roughing pump with oil as indicated in the Operating manual E 750 CAUTION User interface level 1 2 3 4 Setting and maintenance part User part LEVEL This level has very limited information on the alphanumeric display LCD This level is generally selected for production types of applications No access to control keys Cycle key included LEVEL This level allows the operator t...

Page 6: ...STDBY STDBY EVAC EVAC EVAC EVAC TEST TEST SNIFFER SNIFFER SNIFFER SNIFFER CYCLE TEST TEST STDBY STDBY Starting a cycle Ending a cycle Starting Sniffing test mode Ending Sniffing test mode 1 As soon as the inlet pressure reaches 10 mbar 7 5 torr the unit goes in gross leak test mode or when the pressure has reached 5 10 1 mbar 0 37 torr the unit goes in fine leak test mode F1 F3 F2 F4 READY FOR CYC...

Page 7: ...F2 F4 10 mbar hPa Torr 3 10 12 10 11 10 10 10 9 10 8 10 6 10 5 10 4 10 3 10 2 10 2 10 1 1 101 102 103 COR mbar l s STDBY LEAK RATE EVAC TEST Pa m3 s Torr l s 10 7 How to read the He signal analog scale reject point is visualized by a blinking led if the leak value exceeds the reject point the leds will turned red the blinking led will turn orange if the leak value remains under the reject point th...

Page 8: ...um signal zero scale The He signal zero scale displays 2 leds signal centered around the zero value Example 2 leds flash 1 10 7 1 10 8 1 1 10 7 9 10 8 2 10 7 Purpose the zero function offers the operator the possibility to detect small leaks that are smaller than the helium background Activation of zero function connect the part or installation to be tested Operator could find an example in the Op...

Page 9: ...t valve off indicates that the venting valve is not activated closed at the end of cycle The setting by default is off valve closed F1 F3 F2 F4 READY FOR CYCLE INLET VENT off F3 F3 opening closing F3 F3 Opening closing air inlet valve user level 2 3 or 4 To open close air inlet valve with user level 1 C 500 Air inlet vent ...

Page 10: ...uest it can be started by the operator whenever needed the unit has to be off cycle The result of the autocalibration process is displayed External The external autocalibration allows direct readout in cases of operation with an auxiliary pumping system F1 F3 F2 F4 READY FOR CYCLE INLET VENT off F1 F3 F2 F4 READY FOR CYCLE INLET VENT off F1 F3 F2 F4 AUTOCAL IN PROGRESS ELECTR ZERO ADJUST 80 To per...

Page 11: ...t display time of the leak value C 550 To active deactive this function and adjust roughing and measure times C 530 To active deactive this function C 510 To active deactive this function C 560 Assistance to the test Memorization of the latest He signal measured after depressing the CYCLE key at the end of the cycle Device that prevents the unit from getting polluted with Helium Display a greater ...

Reviews: