10
Internal
: The internal autocalibration is automatically activated
during the start-up process. It doesn’t require any
operator action. Thanks to the initial autocalibration,
the leak detector can be immediately operational.
Internal autocalibration on request: it can be started by the
operator whenever needed (the unit has to be off-cycle).
The result of the autocalibration process is displayed.
External:
The external autocalibration allows direct readout in cases of operation with
an auxiliary pumping system.
F1
F3
F2
F4
READY FOR CYCLE
INLET VENT : off
F1
F3
F2
F4
READY FOR CYCLE
INLET VENT : off
F1
F3
F2
F4
AUTOCAL IN PROGRESS
ELECTR. ZERO ADJUST.
80
To perform an external calibration
C 303
AUTOCAL
Calibration